IMP De Mayo Hip Positioner System
OBESE Setup Instructions Download Page 4

De Mayo Hip Positioner

®

 

System

 

OBESE 

(Large Patient)

 Set Up Instructions  

page 4

De Mayo Hip Positioner® is a registered trademark of Innovative Medical Products, Inc.
De Mayo Hip Positioner® US Patent #6,820,621
imp® Rail Clamp is a trademark of Innovative Medical Products, Inc.
© 2011  Innovative Medical Products, Inc.
All Rights Reserved 

OPTION

 - Stabilizer Bar may be inserted into the Anterior Support Assembly with an additional

imp® Rail Clamp (note: maintain 2 fingerbreadths at the rib cage and Horizontal Crossover Bar).

Add the Stabilizing Bar Tube by sliding it from the anterior side of the patient into the opening of 
the Anterior Support Assembly.

Slide the Stabilizing Bar Hook into the imp® Rail Clamp. Attach to the threads of the Stabilizing Bar Tube, 
and tighten both knobs.

In the event (while utilizing the Stabilizing Bar) you
cannot seat the Adjustable Support Assembly,
loosen compression and rotate plate.  With the
Anterior Pelvic Support out of the way, seat the
Adjustable Lateral Support and re-engage the
Anterior Pelvic Support.

Anterior Pelvic Su

pp

pp

pp

pp

ort.

Summary of Contents for De Mayo Hip Positioner System OBESE

Page 1: ...ace then place OR bedsheet on top 2 Pre Assemble the Positioner Assemble flat bar into bottom groove of Lumbar Support Base Assemble washer and locking knob 3 Roll the patient on their side 4 Attach Bariatric Extension Plate 1101 BE 5 Attach Contoured Lumbar Pad for left or right hip application on ...

Page 2: ...e Vertical Support Bar towards the head of the patient Assemble the Horizontal Crossover Large Bar 1 onto the Vertical Bar 1 Allow 2 fingerbreadths between the patient and the bar readths between the patient and the bar Large Anterior Support Assembly includes Anterior Support Anterior Vertical Support Large Anterior Pelvic Support Attach Anterior Rotation Bar 3 to round end of Horizontal Bar 3 2 ...

Page 3: ...rtical Bar 3 and adjust Anterior Pelvic Support 4 so that the Anterior Patient Protector Pad contacts the superior iliac crest st Anterior Pelvic Sup ppo p rt 4 so that the Anterior Pat t 10 Compress Anterior Support Assembly pushing toward Lumbar Support Lock Anterior Support Knob 2 If any readjustments are required release Compression Knob 2 then adjust any other knobs to achieve greatest stabil...

Page 4: ...amp note maintain 2 fingerbreadths at the rib cage and Horizontal Crossover Bar Add the Stabilizing Bar Tube by sliding it from the anterior side of the patient into the opening of the Anterior Support Assembly Slide the Stabilizing Bar Hook into the imp Rail Clamp Attach to the threads of the Stabilizing Bar Tube and tighten both knobs In the event while utilizing the Stabilizing Bar you cannot s...