background image

5. Accessories 

Your Imex LX25P can be used with a number of different 
accessories, including a range tripods, staffs, and wall brackets to 
assist in different applications. 
 
For more information of accessories available please contact your 
local dealer 
 

6. Calibration and Level Checking 

The accuracy of the tool is +/- 1.5mm in 5m. 
Before use check that the instrument is within tolerance. 
 
Place instrument on a solid base i.e. bench top, approx 5m from a 
wall. Turn on horizontal beam.  
At one end of the beam with a pencil mark a fine line on the wall. 
Turn the instrument so that the other end of the beam is at the 
marked line.  
Check difference between current line and marked point, if 
difference exceeds 1mm send back to dealer for calibration. 
 
To check vertical accuracy, utilise a known plumb point such as a 
doorway/ window frame and check that beam lines up. 
 

7. Transportation and Storage 

The laser must be stored in a cool dry environment, free of dust and 
moisture. 
On the completion of use the main On/Off switch must be switched 
to the off position for transportation. 
 

8. Warranty 

The Imex LX25P and LX25P are covered by a 2 year warranty as 
detailed on the warranty form. For registration please visit:  

www.imextools.co.uk

 

 

          Copyright © International Measuring Equipment 2014 

T:

 

+44

 (0)115 9440399 

  

E

:

 

[email protected]

 

  

W:

 

www.imextools.co.uk

 

 

Reviews: