Linearisation with Direct Calibration
If your system is non linear, you can calibrate the meter and correct for this with the
lineariser function.
You will need to apply a series of known loads, starting at 0 and working up to full scale. You
will then tell the meter what it should read for each applied load. You can use up to 10
points.
First, you must select Direct Calibration Mode (see page on calibration Modes)
Then, proceed as shown below...
Lockout Switch must be OFF
Press together, briefly
Set1
Set2
Output
Alarms
Digit
OK
Max/Min
Reset
If the display shows
Lin.OFF
, press Set2
button briefly so that the display shows
Lin.On
, then press OK
Set1
Set2
Output
Alarms
Digit
OK
Max/Min
Reset
1
2
3
Set1
Set2
Output
Alarms
Digit
OK
Max/Min
Reset
The display will show
SEt 00
Ensure no load is applied to the system
and press OK.
The display will now show
rd 00
You can now tell the meter what it should
display for this load.
The display will show
Add.Pt
Press OK
Continued ...
OFF
Circuit board ON
Display will show
Lin.On
or
Lin.OFF
(Default)
Select each digit in turn with the DIGIT button, and increase or decrease
each digit’s value using the UP or DOWN button, until the display is set
as required.
You can set the
Decimal Point
position by pressing the Set 2 button for
3 seconds. All decimal points will illuminate, but one will brighter than
the others. Use the UP/DOWN buttons to choose a decimal point
position and then press OK. Press
4