![Image Engineering iQ-Near Focus User Manual Download Page 8](http://html.mh-extra.com/html/image-engineering/iq-near-focus/iq-near-focus_user-manual_2045174008.webp)
Image Engineering
Seite
8
von
12
3 OPERATING INSTRUCTIONS HARDWARE
3.1 Controls
The following control elements are available on the iQ-Near Focus:
Figure 5: Controls of iQ-Near Focus
1:
Seven-segment display:
Shows the set delay between down/up motions. When the
time has expired, the chart moves back to its starting position.
2:
Red
“Error” LED:
Lights up to alert the user when the device has encountered an
error.
The “Error” LED is operated by the motor control unit.
Examples of possible errors include:
-
The chart could not be slowed down enough and
moved past its END-position.
Solution:
Reset the device by pressing the “Reset” button and perform a homing
procedure by pressing the “Homing” button.
-
The chart is manually forced out of its homing
position.
Solution:
Reset the device by pressing the “Reset” button and perform a homing
procedu
re by pressing the “Homing” button.
If iQ-Near Focus cannot be reset, disconnect the device from the power source,
wait for five seconds, and reconnect it.
If you encounter an error that cannot be resolved, please contact the Image
Engineering Support Team.