3. Power the device up via the start button.
The device will start and, after it has found the recovery key to be present in the serial
port, it will automatically execute the recovery sequence. All viable system data will be
restored and necessary repair steps will be taken without the need of any user interaction.
Note:
The recovery sequence can last some minutes. While the recovery sequence
is running,
no message
will be displayed.
4. When the recovery sequence has finished, the device will power down automatically.
Important:
Do not switch off the device at any time during the recovery procedure!
5. Unplug the recovery key after the device has powered down.
6. Set the compensator into its basic position and insert the compensator plate.
7. Power up the device and launch the scan application in your browser.
The IP address of the device will have the factory default value: 192.168.1.50
8. Change the network parameters to the values which were used before running the
recovery sequence.
Select Setup Device
Poweruser . Locate the section
S2N Base Settings
and
click the button Network . Enter the values for IP address, subnet mask, and default
gateway.
Click the Apply button. Confirm the following message by clicking the OK button.
In the next screen, click the Reboot button.
Reconnect to the device using the new IP address.
9. Select Setup Device
Poweruser . Locate the section
Updates & Uploads
and
click the button Update Firmware . Perform a firmware update.
10. After the firmware update, all software adjustments for the device must be performed.
Select Setup Device
Poweruser . Locate the section
Adjustments & Support
and click the button Adjustments . Perform the adjustments by clicking the
appropriate buttons.
Page 64
Setup and Assembly Manual