The ENABle:OUTOFF command sets the outoff register for the 8033. This register is a filter which
allows certain conditions to turn off the laser output.
Parameters
An <nrf value> whose sum represents the enabled bits below:
1
Interlock broken
2
Misc. hardware errors
Notes
The enabled Outoff bits can be read by using the ENABle:OUTOFF? query. The default value for
this register is three.
Examples
"
ENAB:OUTOFF 1
"action: allows the laser output to be turned off if the interlock is broken.
The ENABle:OUTOFF? query returns the value of the 8033 outoff register.
Parameters
None.
Response
The response data is the binary sum of the following:
1
Interlock broken
2
Misc. hardware errors
Notes
The enabled events or conditions which are used to turn the laser output off can be set by using the
ENABle:OUTOFF command.
Examples
"
ENAB:OUTOFF?
"response: 0, means there are no conditions that will automatically turn off the
laser output.
The ERRors? query returns a list of command, execution, and device errors which have occurred
since the last ERRors? query. These errors are indicated by a number which corresponds to the
type of error which occurred. The error numbers are decoded in Appendix A.
Parameters
None.
Response
Each <response data> consists of an error code value.
Notes
The response data will be a list of the current errors. The errors are represented by numbers and
are separated by commas. A response of zero indicates that there are no errors to report. The
response is sent as character data.
When the error queue is active (errors exist), bit #7 of the Status Byte is set to one.
Examples
"
ERR?
"
response: 0, means there are no errors to report.
"
ERRORS?
"
response: 201, means that the <PROGRAM DATA> value is out of range.
ENABle:OUTOFF
C
OMMON
D
EVICE
D
EPENDENT
F
RONT
P
ANEL
ENABle:OUTOFF?
C
OMMON
D
EVICE
D
EPENDENT
F
RONT
P
ANEL
ERRors?
C
OMMON
D
EVICE
D
EPENDENT
F
RONT
P
ANEL
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for MPS-8033 Series
Page 3: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 9: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 11: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 13: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 23: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 53: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 103: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 107: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...