G E N E R A L I N F O R M A T I O N
Product Overview
2
LDX-3412
C H A P T E R
1
Specifications
Your Comments
Our goal is to make the best laser diode instrumentation available anywhere. To
achieve this, we need your ideas and comments on ways we can improve our
products. We invite you to contact us at any time with your suggestions.
OUTPUT
1
1.
The instrument contains a circuit that turns off the output if a transient is detected.
Current Range:
0 to 200 mA, floating
Compliance Voltage:
>6 Volts
Temperature coefficient:
<100 ppm/
o
C, ambient
Stability, 10-30 min.:
<50 ppm
Noise and ripple (at 100 mA):
<10
µ
A rms
Transients:
<100
µ
A
DISPLAY
Type:
3-1/2 digit green LED
Max Reading:
199.9 mA
Accuracy at 25 ×C:
+0.2 mA
CURRENT LIMIT
Range:
1-200 mA
Accuracy:
+3 mA
PHOTODIODE FEEDBACK
Input type:
Current input from external photodiode
Range:
20
µ
A to 2 mA
Stability:
+0.1%
GENERAL
AC Power:
100, 120, 220, 240 nominal line voltage, ±10%, 50-60 Hz
Operating Temperature:
0
o
C to 40
o
C
Storage Temperature:
-40
o
C to 70
o
C
Warm-up:
1 hour for rated accuracy
Laser Output Connector:
9-pin D-sub
Chassis Ground:
Standard banana jack
Overall Size:
267 x 140 x 66 mm (10.5 x 5.5 x 2.6 inches)
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com