GIGATESTpro
- 7 -
3. MEASUREMENTS
3.1. Turning the instrument on and off, standby, auto power off
Hold the
START
key pressed until the device turns on.
The instrument is turned off after two short pressing/releasing the
START
button,
while no voltage can be applied on the test tips.
The instrument enters standby mode (reduced display brightness) after short
time of inactivity (no key pressed, no voltage applied on the test tips).
From standby mode (to full display brightness), the instrument enters after
pressing any button or by applying the voltage on the test tips.
Auto power off occurs when the instrument is idle (no key pressed, no voltage
applied on the test tips) for about a minute.
After turning off the device can be switched on again after about 1s.
3.2. Notes and principles applicable to all measurements
•
Select required parameter or function by the
R
ISO
/U
SPD
,
FUNC
,
∆
U
TEST
and
∇
U
TEST
keys. The
START
key starts measurement. All set parameters
and functions remain valid until they are changed.
•
If voltage applied on the test tips is > about 10 V, its value is displayed in the
U
IN
area and simultaneously warning symbol “!” is displayed. The
START
button does not start measurement:
Fig. 3.1
Example of voltage measurement
•
If battery is low (only red part of battery indicator is displayed), then you can’t
start the measurement by the
START
key – after pressing it the low battery
symbol is displayed for a while. Thereafter, the instrument goes into status
before pressing the
START
key. Battery must be replaced / accumulators
charged as described in the Chapter 4.1.
Fig. 3.2a
Indication of low battery
Fig. 3.2b
Low battery after the
START
key was pressed