background image

 

 

 

 

 

Page 1 of 1 

21/03/200716:27

 

December 2006 

 

 

USER INSTRUCTIONS 

MG500P

 

ILFORD MULTIGRADE 500P EXPOSURE PROBE 

FOR USE WITH THE ILFORD MULTIGRADE 500 EXPOSING SYSTEM

 

The 

ILFORD MULTIGRADE 500P

 exposure probe is for use with the ILFORD MULTIGRADE 500 exposing system, to 

provide automatic assessment of exposure time at the grade selected from a spot reading. 

 

1 CALIBRATION

 

The calibration system is very straightforward, and provides a sound basis for obtaining correct exposure times when using 

the probe. The probe is supplied with a test negative to help with initial calibration. When you become more familiar with 

the probe you may wish to carry out the following procedure using a negative that you consider to be more representative 

of your own work. 

 

a) Using the test negative supplied, make a satisfactory print on ILFORD MULTIGRADE paper using, for example, grade 2. 

Make a note of the exposure time and grade used. The exposure time and the enlarger settings must not be changed during 

the remainder of the calibration procedure. 

 

b) Plug the MULTIGRADE 500P exposure probe into the ‘probe’ socket at the rear of the control unit. 

 

c) Press ‘focus’ (MULTIGRADE 500CPM control units) or ‘auto’ (MULTIGRADE 500C control units.) 

 

d) Position the probe photocell on the projected image on the area bounded by the clear ring. Once the probe is 

positioned, do not move the probe or lean over it while the probe is taking a reading. 

 

e) Press and release the probe switch. The probe provides information to the control unit which, in turn, calculates and 

displays the optimum exposure time, based on image intensity. The calculation incorporates a correction factor to take into 

account any background illumination, e.g. safelighting. 

 

f) If the display does not agree with the exposure time noted in operation (a), adjust the probe calibration knob to a lower 

setting (towards 0) if a shorter exposure time is displayed, or to a higher setting (towards 9) if a longer exposure time is 

displayed. 

 

g) Repeat operations (e) and (f) until the display agrees with the exposure time noted in operation (a). 

Note. In all cases, choose the calibration knob setting that gives the closest match. Never attempt to position the knob 

between numbers. 

 

h) If calibration cannot be achieved within the range of the knob, carry out the procedure detailed in section 2 below. 

Reviews: