15
• Gamma Control: Used to adjust the gamma level from 1.8 to 2.6 in 0.2 increments.
D14: Gamma level at DICOM Part 14.
User: Selectable from the user-preset gamma levels (factory option
setting).
• Signal Set.:
Used to set the format of each input signal. RGB or YPbPr can be
set.
5.9 Memory Select Screen
This screen is used to make the user memory settings and to return them to the factory
ones.
Display screen
Status
User
Display
Image
Color
Memory
User Memory
User1/User2/User3/User4/User5/User6/User7/User8
Cancel/User/Display/Color/ALL
User1
Factory Reset
Cancel
Input Control
Video
S-Video
Component
PC
DVI
SDI
○
○
○
○
○
○
×
×
×
×
×
×
IN1
IN2
Switch behavior description
IN1: CH name - Signal name
IN2: CH name - Signal name
Selectable
:○
Not selectable
:×
OP not mounted
:N/A
Memory select screen description
• User Memory: Used to memorize up to eight different groups of settings.
All the settings made on the menu screens can be put in memory.
• Input Control: Used to set the inputs selectable for Input1 and Input2.
With the “
○
” sign, the input is selectable. With the “
×
” sign, it is
not selectable.
When a relevant option is not mounted, the inputs are expressed
with N/A.
• Factory Reset: Used to return all the modified settings to the factory ones.