17
092007
Any malfunctions during operation will be identified by an error message on the display (E and F).
Proceed as follows in such cases:
!
Switch off device using the main switch (A)
!
Carry out corrective measures
!
Restart the device
If the actions described fail to resolve the fault or another error code is displayed then take one of the following steps:
- Contact the service department
- Send the device for repair, including a short description of the fault.
Error codes
Error code
Cause
Effect
Solution
Undervoltage
Err
Mains voltage failed during operation.
Device off
- Switch device off and on
Err
Mains plug connected with switch set to
Device off
- Switch device off and on
“On”.
Overtemperature
Err
Device will switch automatically to cooling Motor speed - Allow device to cool down:
mode if operated with excess load
set auto-
- Allow to continue running in cooling mode
matically
- Switch device off. Switch on again after some time.
(not adjust-
(Duration depends on the magnitude of the overtempertature)
able)
A standard power meter is used between the mains socket
and the mains plug to prevent the device failing as a result
of excess load.
Power consumption < 500 W : continuous operation
Power consumption > 500 W : device may switch to cooling mode
Electronic
Err
Electronic fault
Motor speed - Switch off device
set auto-
Please send the device to the Servicing department for repair,
matically (not enclosing a short description of the fault.
adjustable)