Auto Zeroing:

For compensation mechanical friction and grease viscosity, you must auto-zero the device before you measure the viscosity:
Before auto-zeroing, empty the vessel according to the reminding of the 

WiCo

 firstly.

Set the maximum speed higher a little than the select speed in the following viscosity measuring.
For example: when you use R 1345 HV and the optimum speed (200 rpm) to measure a medium viscosity, the maximum speed should 
be set as about 210 to 250 rpm.
The zeroing processing will take about 150 seconds.

Viscosity measuring:

In order to achieve optimum measurement result, the device should run idle (without stirrer tool) more than 30 minutes at the speed 2000 
rpm. Then, perform auto zeroing as above instructions.

After auto zeroing, you can start to measure the medium viscosity as following: 

)

 Input the density for the test medium.

)

 Select the stirrer tool as required. 

)

 Active the viscosity display. 

)

 Back to working screen and set speed for measuring.

)

 Start to run the device.

Ö

 Read the viscosity value when the figure get steady.

Note:

 

The formulas for calculating the viscosity have been put together for the respective viscosity range with the vessels listed and the 

stirring elements appertaining thereto, based upon the physical laws and taking into account the smallest error square out of more than 
20 measuring values (with Newtonian calibration oils at 22 ºC).

Measuring error calculating formula:

Example for viscosity measuring:

When you measure a medium in following status:
Kinematic viscosity range: 10000 mpa

·

s to 100000 mpa

·

Medium density: 0.980 g/

cm

3

Stirrer tool: R 1302
1. Warm up the device:

Remove the stirrer tool from the device and switch the device to  range II (high speed range). Run the device with speed 2000 rpm 
more than 30 minutes. 

2. Perform auto Zeroing to the device:
  Stop the device and enter the menu option "Auto Zeroing" under "Viscosity".
  Press on "OK" in "Empty the vessel" screen.
  Set the maximum speed (e.g.:210 rpm) to the auto zeroing. The maximum speed should be higher a little than the speed that will be 

used for viscosity measuring in next step.

  Press on "OK" to start the auto zeroing. This process will take about 150 seconds.

Menü

Viskosität

Viskosität

Dichte

R

ührer

Display

Automat. Nullabgleich

> Viskosität

Automat. Nullabgleich

Gefäß leeren!

g

OK

> Viskosität

Automat. Nullabgleich

OK

Max. Drehzahl:

       50   

rpm

> Viskosität

Automat. Nullabgleich

OK

Max. Drehzahl:

      210  

rpm

> Viskosität

Automat. Nullabgleich

Bitte warten...

> Viskosität

Automat. Nullabgleich

OK

Max. Drehzahl:

      210  

rpm

B - A

A

= X %

A – Standard oil viscosity
B – Measured viscosity 

Summary of Contents for STARVISC 200-2.5 control

Page 1: ...20000015716 IKA STARVISC 200 2 5 control STARVISC 200 2 5 control_062018...

Page 2: ...e setup Dangerous parts Fig 3 1 2 3 4 Fig 2 Fig 4 H I J G G K L M Fig 1 A J I H G F E STARVISC 200 2 5 control station Wireless Controller WiCo Q B M O N C L K P D Fig 7 Fig 5 Fig 6 T U Q S U X I G Z...

Page 3: ...he equipment This device complies with Part 15 of the FCC rules subject to the following two conditions 1 This device may not cause harmful interference and 2 This device must accept all interference...

Page 4: ...r safety can not be ensured simply with design requirements on the part of the product For this reason it may become necessary for users to take other precautionary safety measures For example glass d...

Page 5: ...is fully discharged the only means of switching the station off are the safe STOP and the off switch Please note the following safety in structions for the battery pack RB 1 re chargeable battery Kee...

Page 6: ...ce is operated improperly or contrary to the manufac ture s speci cations if the device or the printed circuit board are modi ed by third parties Drive The rotary knob B see Fig 1 of the WiCo allows t...

Page 7: ...verhead stirrer must be secured to a stable stand e g R 2722 or R 2723 with a cross sleeve e g R 270 The stirring vessel must always be securely xed for safety reasons You must also ensure that the mo...

Page 8: ...emperature sensor cable should be plugged into the appropriate port after the covers have been removed as shown in Fig 6 Once the device has been connected to the PC using the USB data cable it will t...

Page 9: ...to search for the WiCo A beep signal will be audible even if the WiCo is switched off Speed setting You can use the rotary knob B see Fig 1 to pre set the required speed before you start the device I...

Page 10: ...station is communicating via a USB cable The symbol no longer appears if no USB cable is being used for communicating with the station Home This symbol means that the WiCo is connected to the station...

Page 11: ...the Menu button N Select the menu by turning the rotary push knob B to the right or left to select the desired menu or sub menu which can then be selected by pressing the rotary push knob Press or tu...

Page 12: ...ate Info Black White Standard Mode Battery Mode Volume Key Tone Device Name Remote Control Bluetooth WiCo Other Stirring Safety Software Max Speed Max Torque Interval Run Time Out Station Version Inte...

Page 13: ...for stirring tasks which require metering and feeding additives Torque Display The Torque Display menu allows the user to specify that the torque is shown in the display A tick shows that the option...

Page 14: ...suring error max 12 Sample volume 1500 ml Necessary beaker HF 3000 DIN 12332 3000 ml low shape 1500 ml 150 mm 210 mm 10 mm R 1302 Measuring condition Paddle stirrer R 1375 Measuring condition Anchor s...

Page 15: ...nd the stirring elements appertaining thereto based upon the physical laws and taking into account the smallest error square out of more than 20 measuring values with Newtonian calibration oils at 22...

Page 16: ...mer is interrupted 4 Select the stirrer tool Select the proper the stirrer tool R 1302 HV Press on OK to confirm the setting Viskosit t R hrer OK R 1345 HV Viskosit t R hrer OK R 1302 HV Viskosit t R...

Page 17: ...gram segment in the program When user edits the program time for at least one segment tick for respective program will appear Delete Deletes the selected program If a selected program is deleted by pr...

Page 18: ...ode Edit Delete Insert 1 2 3 4 5 150 0 00 10 00 00 Yes No Programs Start Edit Delete Program 1 Program 2 Program 3 Program 4 Program 5 PROGRAM 1 Seg No rpm Time hh mm Interm Mode 1 2 3 4 5 0 00 00 No...

Page 19: ...ou to select the desired lan guage by turning and pressing the rotary push knob B A tick indicates the language that is set for the system Units The Units option allows you to select the desired unit...

Page 20: ...at the computer s request Command syntax and format The following applies to the command set Commands are generally sent from the computer Master to the stirrer machine Slave The stirrer machine sends...

Page 21: ...to the computer A USB micro B USB A 9 pin RS 232 9 pin RS 232 or micro B A USB micro B USB A micro B USB A USB A or A 9 pin RS 232 9 pin RS 232 USB micro B USB A micro B USB A USB 2 0 cable micro A mi...

Page 22: ...ou require servicing return the device in its original packag ing Storage packaging is not sufficient Please also use suitable transport packaging Warranty In accordance with IKA warranty conditions t...

Page 23: ...tirrer n 1000 R 1342 Propeller stirrer 2000 R 1345 Propeller stirrer n 800 R 1381 Propeller stirrer 2000 R 1382 Propeller stirrer 2000 R 1385 Propeller stirrer 800 R 1388 Propeller stirrer 400 R 1389...

Page 24: ...e VAC 230 10 115 10 100 10 Frequency Hz 50 60 Max input power W 130 Max output power at stirring shaft W 84 IP code according to EN 60529 IP 40 Protection class I Excess voltage category II Contaminat...

Page 25: ...W x D x H mm 71 x 74 x 151 Weight kg 0 28 Ambient temperature C 5 40 Ambient humidity relative 80 IP code according to EN 60529 IP 40 Interface USB Battery pack Voltage V 3 7 Charging capacity mAh 20...

Reviews: