• Place the RV 10.8001 condenser seal
(U)
in the condenser
receptacle and fit the glassware to the device according to the
assembly instructions. (Fig. 16, 17 and 18)
Mount the condenser
• Fix the receiving flask and hose connectors as shown. See also
operating manual of glassware. (Fig.18)
Fig.17
Fig.18
W
V
U
Fig.16
• Place the cap nut into the condenser, and then put the spring
ring to the flange of the condenser (
V
).
• Place the condenser on the rotary drive (
W
) and tighten the cap
nut by hand. (Fig.17)
Removing the condenser
• Use the ring spanner provided to loosen union nuts that are
tightly fitted.
• Loosen the union nut by turning anticlockwise.
Description of special condensers
(see products/accessories at
www.ika.com
)
•
RV 10.3 Vertical-intensive condenser with manifold
This vertical-intensive condenser features a double jacket design
for particularly efficient condensation.
Also available with coating (RV 10.30)
•
RV 10.4 Dry ice condenser
Dry ice condenser for distilling low-boiling solvents. Cooling
by dry ice, no cooling water required. Maximum condensation
thanks to low temperatures.
Also available with coating (RV 10.40)
Mounting the washer
Drainage
washer
Note: Pay attention to the
correct position of the washer.
Fig.19b
Fig.19a