36
Glassware
Fit the condenser seal
• Place the condenser seal in the condenser receptacle and fit the
glassware to the device according to the assembly instructions.
Heating bath
Refer to the chapter “Commissioning“
in the heating bath instruction manual!
• Place the heating bath on the stand of the rotation drive and
push it into the left position.
Note:
Data is exchanged between the drive unit and the heating
bath by means of an infrared link (
1
). Please note that reliable
communication is only guaranteed when the infrared beam has
clear line-of-sight to the detector.
CAUTION
IR
1
NOTE:
Please read the operating instructions of the glassware for
the safe handling of laboratory glassware !
• There is a locking knob for locking or unlocking the locking
device.
If a red mark is shown, the locking device is unlocked, otherwise
it is locked.
To lock or unlock the locking device, push the locking knob to
the end position.
• Open the locking device
on the drive head by turning it 60°
counterclockwise. Then you can see a red mark .
• Feed the vapour tube in until it stops.
• Then, lock the locking device by turning it clockwise by 60°.
• Pushing the lock knob to the end position, the red mark should
be covered and invisible.
• The vapour tube is not allowed to be pulled out!
• Check the correct axial locking device on the vapour tube.
• Keep the red mark invisible.
CLICK
Summary of Contents for RV 10 auto FLEX
Page 2: ......
Page 131: ...131 25 C 1...
Page 135: ...135 30 1 30 RV 10 auto 1 140 2 1 2 1...
Page 136: ...136 1 IR 1 60 60 CLICK...
Page 137: ...137 1 2 4 1 3 5 6 4 6 1 2 5 6 4 3 1 2 3b 3a 3c 3a 3a 2 3b 20 120 3b 1 2 3a 3b 3c...
Page 140: ...140 RV 10 5001 RV 10 5001 1 2 1 2 RV 10 5001 RV 10 4002 RV 10 control LOAD...
Page 149: ...149 2 3 1 2 1 0 6 3 HB 2 3 2 3 IKA 50 3 IKA RV 10 3000 150 1 2...
Page 152: ...152 RV 10 4002 RV 10 4002 A...
Page 153: ...153 40 C RS 232 USB LWS RV 10 RV 10 LWS IKA 24...