IKA RO 10 Manual Download Page 5

12

Pressing the +/- buttons 

(H

 or 

I)

 increases or reduces the 

temperature.

RT 

 

The heating function is started and stopped by pressing the 
„Temp“ button

 (J)

 Function Heating

 ECO mode

RO/RT

The heat generated in the device can be reduced by lowering the 
power of the input coils.
Pressing the “ECO Mode“ 

(F)

 key to reduce the current power 

setting. ECO mode is indicated by an “E” preceding the speed. 
When the power is set to ECO mode, the speed is limited to 600 
rpm.

To make the stirring power more independent from the speed 
range, press the “ECO Mode“ membrane key (

F

) and then select 

one of the following setting options.
P50  (Stirring power 50% independent of the speed range)
P75  (Stirring power 75% independent of the speed range)
P100  (Stirring power 100% independent of the speed range)
----   (Standard stirring performance with the stirring power 
 

regulated by the speed)

RO/RT 

 

The 

(G)

 key activates the “Reverse Rotation” function. 

When the „Reverse Rotation“ function is switched on, a decimal 
point 

(K)

 appears in the LCD display. 

 

The running time (30, 60, 90 or 120sec) is set by pressing the 

(G)

 

key followed by the +/- keys. 

 

After the preset running time, the drive is switched off and 
restarted at minimum speed in the opposite direction after a 
pause specific to the speed. 

In this way the magnetic rod is captured and any liquid still 
rotating is gradually slowed down. Then the drive accelerates to 
the target speed. 

 

Note: If the “Reverse Rotation” key 

(G)

 is pressed again, the drive 

will keep rotating in the same direction. 

 

When the device is started up again, the drive will rotate in the 
direction last set.

 Direction of rotation reversal

RO/RT 

 

To toggle between modes, press the Start/Stop key (C) when 
switching on. 

 

Operating Mode “A” 
All the parameter settings are retained when the device is 
switched off or disconnected from the mains. 
The “Start/Stop” 

(C)

 key starts the device with the set parameters. 

 

Operating Mode “B“ 
All parameter settings, including “Start/Stop”, are saved when the 
device is switched off. 

 

Operating Mode “D“
In the D mode the device operates as per the A mode but with the 
following exception:
- Switch on and then press the “Temp” membrane key to confirm 

the safety temperature. 

- The value flashes until it is confirmed.

Factory setting: mode A

 Operating modes

The equipment is maintenance-free. It is only subject to the natu-
ral wear and tear of components and their statistical failure rate.

Cleaning

   

For cleaning disconnect the main plug..

Use only cleansing agents which have been approved by 

IKA

®

 to 

clean devices.

 Maintenance and cleaning

Use to remove:

Dyes Isopropyl 

alcohol

Construction materials  water containing ten side / isopropyl alcohol
Cosmetics 

water containing ten side / isopropyl alcohol

Foodstuffs 

water containing tenside

Fuels 

water containing tenside

-  Do not allow moisture to get into the appliance when cleaning.
-  Wear protective gloves when cleaning the devices.
-  Please consult with 

IKA

®

 before using any cleaning or deconta 

  mination methods, other than those recommended here.

Summary of Contents for RO 10

Page 1: ...IKA RO 5 10 15 RO RT_072017 IKA RT 5 10 15 20000004954 ...

Page 2: ...ccurate results ATTENTION Risk of damage due to magnetism DANGER Reference to the endangerment by a hot surface EN General hazard ATTENTION Contents Page Control panel and display 02 Contents 09 Declaration of conformity 09 Warranty 09 Explication of warning symbols 09 Safety instructions 10 Unpacking 11 Correct use 11 Commissioning 11 Safe temperature limit 11 DANGER Setting the safe temperature ...

Page 3: ...he drive mag nets at high motor speeds even if the heater is not operational ATTENTION DANGER WARNING Process pathogenic materials only in closed vessels under a suitable extractor hood Please contact IKA if you have any questions Do not operate the appliance in explo sive atmospheres with hazardous substances or under water Only process media that will not react dangerously to the extra energy pr...

Page 4: ...ECO mode is active If the drive has not been started the display reads OFF If the drive has been started the selected speed is displayed Commissioning RO RT Observe the ambient conditions temperature humidity etc listed under Technical Data The device is switched on and off via the switch A located on the right hand side of the device On the front panel a small arrow labelled Power above the displ...

Page 5: ... Note If the Reverse Rotation key G is pressed again the drive will keep rotating in the same direction When the device is started up again the drive will rotate in the direction last set Direction of rotation reversal RO RT To toggle between modes press the Start Stop key C when switching on Operating Mode A All the parameter settings are retained when the device is switched off or disconnected f...

Page 6: ...ature safety tempe Carry out an internal test on the device to check the plug in connector rature 40 K for the temperature sensor Er31 Triac error detection doesn t work Heating off Switch off device Warning Only to be carried out by authorised service personnel Check PCB Er46 Difference between temperature of safety Heating off Switch off device sensor and temperature of control sensor Warning On...

Page 7: ...95 Dimension mm 120 x 610 x 60 190 x 610 x 60 280 x 610 x 60 Weight kg 4 0 6 5 9 4 Adjustment and display resolution K 1 Surface temperature C 120 Safety temperature limit adjustable C 50 150 RO RT 5 RO RT 10 RO RT 15 Fuse 230 V A T 2 5x20 T 4 5x20 T 6 3 5x20 100 V 115 V A T 4 5x20 T 6 3 5x20 T 10 5x20 RO RT Setting resolution rpm 5 Deviation between the stirrer points 0 Permissible duration of op...

Reviews: