Menu settings

Menu structure:

Menu

Mode

Alarm and key tone

Factory reset

Info calibration

 

T mode

A
B
C

On
Off

No
Yes

Edit (Edit CALI) 

On
Off

Calibration and adjustment

 

C0
C1
C2
C3

Factory setting

activated
-
-

activated
-

activated
-

activated
-
-
-

-
activated

 Info calibration:

In this menu, the current offset values can be viewed�

 Edit calibration:

In this menu, the offset values can be entered directly�

 Calibration and adjustment:

This menu shows how offset values can be entered after the respective temperatures are approached and with 
an external measuring device cross-checked (calibrated)�

Navigation in the menu:

F

  Enter/leave the menu by pressing ”Counter/Timer” key (E) and “Timer“ key (H)�

F

  Press the“Timer (+)“ key (K) or “Timer (-)“ key (L) to switch the menu options�

F

  Change the menu setting by pressing “Temp (+)“ key (O) or “Temp (-)“ key (P)�

F

  Save the setting by pressing the “Start/Stop“ key (I)�

Locking the control elements:

The control elements of the device allow to be locked by 
pressing the lock key (G) about 2 seconds, so no accidental 
changes during operation are possible (LED (F) lights up)�
By pressing the lock key (G) about 2 seconds again, the con-
trol elements are released (LED (F) no longer lights up)�

Switching between “Counter” and “Timer” function:
The “Counter” function could be switch to “Timer“ function 
by pressing the “Counter/Timer” key (E)”� The timer indica-
tor (J) lights up to indicate the “Timer“ function is activated�
Press the “Counter/Timer” key (E) again, the “Counter“ func-
tion will be activated and counter indicator (D) lights up�

Timer function:
The “Timer” can be adjusted by pressing “Timer (+)” key (K) 
or “Time (-)” key (L)�
During operating, the “Timer” start increase with hour/min-
ute (hh:mm) mode�
When the timer value reach 24 hours, the display switch 
to day mode� The display show the day and hour/minute 
(hh:mm) alternatively�
For changing the timer more quickly, Keep pressing on “Tim-
er (+)” key (K) or “Timer (-)” key (L)� 
After setting, the timer can be started by pressing “Timer” 
key (H)� The heating function stops automatically once this 
time has expired� The display (C) blinks “00:00” until you 
press any key�

Note: 

The maximum timer value is 100 days (d: 99/23:59)� 

The timer value could be changed in standby or operation 
process� When timer function is activated, starting or clos-
ing the heating function by pressing “Start/Stop“ key (I) 
doesn’t influence of the timer function�

Summary of Contents for Oven 125 basic dry

Page 1: ...Operating instructions EN OVEN 125 basic dry OVEN 125 basic dry glass...

Page 2: ...ion 1 Mains switch 2 Door handle 3 Device foot 4 Operator Panel and display 5 USB interface and temperature limit 6 Mains socket 7 Venting flap Device setup Dangerous parts Dangerous parts Device setu...

Page 3: ...bols 4 Safety instructions 4 Intended use 6 Useful information 6 Transportation and storage 7 Unpacking 7 Installation 7 Operator panel and display 11 Setting the temperature limit 12 Commissioning 12...

Page 4: ...an be accessed by everyone Ensure that only trained staff work with the device Follow the safety instructions guidelines occupational health and safety and accident prevention regulations Transport an...

Page 5: ...The voltage stated on the type plate must correspond to the mains voltage The device can only be disconnected by pulling out the mains plug Socket must be earthed The device will automatically restar...

Page 6: ...libration and offset factory calibration CAL 1 5K CAL 3 5K CAL 2 7K resulting offset graph offset 10 K deviation between offset points 30 K Individual load and ambient condition might influent the tem...

Page 7: ...and corrosive ambient air during transportation and storage It is also important to ensure that the relative humidity does not exceed 80 The device must be completely emptied before transporta tion an...

Page 8: ...l The screws used for fixing is selected according to the wall conditions and is not included As indicated in following figures attach the device to a wall with the tilt protections included with the...

Page 9: ...device Never stack more than two devices When stacking the device four persons are required to lift it In order to keep enough space please use stacking adapter recommend by IKA 150 mm 150 mm Stacking...

Page 10: ...avy about 15 kg of the door Open the door Hold the door and loosen the hinge pin with a suitable tool Remove the door carefully Alternate the position of the keepers and hinges with 4 mm hexagonal wre...

Page 11: ...Timer Indicate the timer function is activated K Timer key Increase the timer setting value Navigation selecting the settings in the menu L Timer key Decrease the timer setting value Navigation selec...

Page 12: ...function is activated by pressing the Start Stop key I the heating indicator T lights up Then the target temperature and the actual temperature appears on display A alternately Accordingly setting val...

Page 13: ...hts up By pressing the lock key G about 2 seconds again the con trol elements are released LED F no longer lights up Switching between Counter and Timer function The Counter function could be switch t...

Page 14: ...result This function allows the temperature to be calibrated and adjusted to three dif ferent temperature values When calibration locate a calibrated reference measuring sensor at the center of the c...

Page 15: ...t side of the screen and actual temperature appears on the left side of the screen Wait about 1 2 hours until temperature is stable e g reach about 98 8 C and Set LED S no longer flashes and the follo...

Page 16: ...ndations for the design of electrical plug connections for analogue and digital signal transmission on individual items of laboratory control equipment rev 1 1 The NAMUR commands and the additional sp...

Page 17: ...Maintenance and cleaning The device is maintenance free It is only subject to the natural wear and tear of components and their statistical failure rate Cleaning For cleaning disconnect the mains plug...

Page 18: ...No internal sensor Heating off Contact IKA service department If the actions described fails to resolve the fault or another error code is displayed then take one of the following steps Contact the se...

Page 19: ...display LED Max timer counter 100 days d 99 23 59 hh mm Max number of tray 6 Max load for one tray kg 30 Interface USB Permissible ambient temperature C 5 40 Permitted ambient humidity 80 Permitted o...

Page 20: ...s lab ika my CHINA IKA Works Guangzhou Phone 86 20 8222 6771 eMail info ika cn POLAND IKA Poland Sp z o o Phone 48 22 201 99 79 eMail sales poland ika com USA IKA Works Inc Phone 1 910 452 7059 eMail...

Reviews: