25
KS 4000 i 1208
Error code Description
Cause
Effect
Solution
Er 3
Temperature inside unit too high
• Permitted ambient temperature has
Heating off
• Switch off the unit. Allow it to cool down
been exceeded
and then switch on again.
• Ventilation slots or fan housing
• Clean ventilation slots or fan
blocked
housing
• Observe maximum permissible
ambient temperature
Er 4
Differanceberween setpoint and
• Motor seized or overloaded
Heating off
• Reduce the torque load
actual speeds is too large
(load)
• Reduce the setpoint speed
Er 8
The calibration value of the tempe- • Fault in calibration procedure
Heating off
• Repeat the calibration procedure
rature sensor is outside the limit
• Value was incorrectly stored to memory
value
• EPROM switching error
Er 14
External temperature sensor,
• Short-circuit at temperature sensor
Heating off
• Check the plug
short-circuit
plug
• Substitude the temperature sensor
• Short-circuit in connecting cable
or on temperature sensor
Er 16
External temperature sensor has
• SAFE Temp is set at a temperature
Heating off
• Allow the unit to cool
exceeded the SAFE Temp
lower than the actual temperature on
• Set the SAFE Temp at a higher temperature
the external temperature sensor
Er 17
Temperature sensor - Incubation-
• SAFE Temp is set at a temperature
Heating off
• Allow the unit to cool
room has exceeded the
lower than the actual temperature on
• Set the SAFE Temp at a higher temperature
SAFE Temp
the external temperature sensor
Er 26
Difference between the internal
• Ventilation slots in incubation room
Heating off
• Switch off the unit. Allow it to cool down
contro and safety temperature
blocked
and then switch on again.
sensors too large
• Radial-flow fan does not rotaded
• Check fan or ventilation slots and
• Fault in the control or safety
clean if necessary
temperature sensors
Er 50
Appliance imbalance
• Proper motion of appliance too
The target speed was
• Select a lower speed setpoint
heavy
reduced several times
• Distribute load evely
• Load is asymmetrical, has slipped or
automatically
• Reduce load
is too heavy
Er 60
Power outage
• Power outage during operation
Interruption of heating
• Delete the display by pressing the
or shaking function
shift key
3729600a_KS4000_122008:KS 130 b-c 28.01.2009 10:34 Uhr Seite 25