27
Error codes
Any malfunctions during operation will be identified by an error message on the display�
Proceed as follows in such cases:
Switch off device using the main switch at the back of the device
Carry out corrective measures
Restart device
Error code Effect
Cause
Solution
Error 01
Pump off
Heating/refrigerating off
No external sensor
- Check this sensor
Error 02
Pump off
Heating/refrigerating off
Motor over current (rate current)
- Reduce pump motor speed
- Use fluid with lower viscosity
- Check if the pump impeller is blocked
Error 03
Pump off
Heating/refrigerating off
Motor over current (Max current)
- Reduce pump motor speed
- Use fluid with lower viscosity
- Check if the pump impeller is blocked
Error 04
Pump off
Heating/refrigerating off
Motor hall signal missing
- Reduce pump motor speed
- Use fluid with lower viscosity
- Check if the pump impeller is blocked
Error 05
Pump off
Heating/refrigerating off
Too high liquid level
(when the heating function is activated)
- Check the fluid level and buoyage
Error 06
Pump off
Heating/refrigerating off
Too low liquid level
(when the heating function is activated)
- Check the fluid level and buoyage
Error 07
Pump off
Heating/refrigerating off
Too high voltage
- Check the mains power
Error 08
Pump off
Heating/refrigerating off
Too low voltage
- Check the mains power
Error 09
Pump off
Heating/refrigerating off
Device internal temperature is too
high
- Check the ambient temperature and let
the device cool down
- Check the fan and clean the grids at the
rear side
Error 10
Pump off
Heating/refrigerating off
PC communication failure
- Check communication cable
Error 11
Pump off
Heating/refrigerating off
Temperature difference between
control sensor and safety sensor is
too much
- Check safety temperature circuit and
bath fluid
Error 12
Pump off
Heating/refrigerating off
Safety temperature alarm
- Check the bath temperature measurement
Error 13
Pump off
Heating/refrigerating off
Heater switched off by safety circuit - Check safety temperature set value, fluid
level
Error 14
Pump off
Heating/refrigerating off
Fan error
- Restart the device
Error 15
Pump off
Heating/refrigerating off
Safety shutdown of the compressor
- Restart the device
Error 16
Pump off
Heating/refrigerating off
Errors at the compressor (suction
side)
- Restart the device
Error 17
Pump off
Heating/refrigerating off
Errors at the compressor (pressure
side)
- Restart the device
Error 20
Pump off
Heating/refrigerating off
Fan cooling system error
- Restart the device
Error 21
Pump off
Heating/refrigerating off
Internal error
- Restart the device