ONLINE ELECTRONICS LTD
ID5002P MANUAL RevA01
Page 13 of 21
3.6.3.
MAIN HOUSING PIPELINE MOUNTING
The pipeline mounted unit is supplied with a set of ratchet
straps or steel banding as shown.
1.
Ensure the mounting plate is securely fitted to the
housing as per 3.6.1 MAIN HOUSING MOUNTING PLATE.
2.
Temporarily place the unit at the chosen location to
ensure that the unit sits in place without rocking
excessively.
3.
Thread 2x loops of banding through the slots provided in
the mounting plate as shown and then following the
instructions provided with the banding gradually tighten
the banding until the unit is held firmly. Do not use
excessive force.
4.
If equipotential bonding or earthing is required then the
M4 threads around the perimeter of the display endcap may be used.
3.6.4.
SENSOR MOUNTING
The sensor is supplied with a mounting
plate, a mounting stud and a length of
mounting band.
1.
Pass the mounting stud through the
ø6mm hole in the mounting band and
then screw into the mounting plate
approximately 4 full turns. Do not
tighten further at this point.
2.
If the pipeline surface temperature is
expected to exceed +110°C then the
sensor must be fitted on the lower half
of the pipeline as shown to minimise
convection
heating
and
Online
Electronics should be contacted for
further guidance.
3.
Loop the mounting band around the pipeline and then following the instructions provided
with the banding gradually tighten the banding until the mounting assembly is held firmly.
Do not use excessive force.
4.
Ensure that the mounting plate is positioned to form a cross with the mounting band and
the mounting band has been firmly tightened before fully tightening the mounting stud
into the mounting plate using a 21mm AF spanner. It should not be possible to move the
mounting plate at this point.
5.
Attach the ID5002P sensor to the mounting stud tightening firmly using a 1” AF spanner.