LNU sensor : Installation & programming
Installation Guide
K0J00016-EN-A02
IJINUS
–
25 ZA de Kervidanou 3
–
29300 MELLAC
–
FRANCE
-
7
b.
First step : Run Avelour and connect to the sensor to configure
After connection of the Wiji kit equipped with its antenna (or the Wiji key) on your laptop or PC USB port, run
the software Avelour version 6.04 minimum. Click on the button “Connect to a wireless device” then the sensor
or logger will be directly visible by its part number (SN) without the need to activate anything else (ex: IJA0101-
0000 3559). Locate the sensor's serial number (SN) on the sensor label and click "OK".
On the first connection with the sensor only the serial number will appear. The following connections you will
see also the name of the installation site or another name that you define.
c.
Second step : Selection of the sensor to configure, eventually a firmware update would
be suggested
During its connection with the sensor, the following pop up appears:
When connected, and only if the sensor already has measures in memory, the following options will suggest you
to retrieve the measures:
After your choice of these options, if the sensor is not up to date (case of new firmware developed since your
last connection or you are now connected with a new version of the software Avelour). It is strongly
recommended to read carefully the different messages in the options windows.
The firmware update can last a moment so it is advised to do it while in your office. On-site prefer the best radio
connection (so not if the metal cover of a man hole is closed).