Problem
LCD shows low battery
symbol
LCD shows “Er 0”
LCD shows “Er 1”
LCD shows “Er 2”
LCD shows “Er 3”
LCD shows “Er 4”
LCD shows “Er 5”
LCD shows “Er 6”
LCD shows “Er 7”
LCD shows “Er 8”
LCD shows “Er A”
No response when you press
button or install batterypiles
Low Battery
Pressure system is unstable before
measuremente
Pneumatic system blocked or cuff is too
tight during inflation
More than 3 minutes with cuff pressure
above 15 mmHg
Incorrect operation or strong
electromagnetic interference.
EEPROM accessing error
Pneumatic system leakage or cuff is too
loose during inflation
Fail to detect systolic pressure
Cuff pressure above 300mmHg
Device parameter checking error
Pressure sensor parameter erroron
Fail to detect diastolic pressure
Change all the batteries
Try again without moving
Take out the batteries, wait
five minutes and then
reinstall batteries
Apply the cuff correctly
and try again
Measure again after 5
minutes. If the error is still
there, please contact your
local iHealth distributor
Explanation
Solution
9