iHealth KD-557BT Owner'S Manual Download Page 6

 

 

 5 

 

 16

Note:  

a.Please refer to the cuff circumference range in “SPECIFICATIONS” to make sure that the 

appropriate cuff is used. 

b.Measure on the same arm each time. 

c.Do not move your arm, body, or the monitor and do not move the rubber tube during measurement. 

d.Stay still, calm for 5 minutes before blood pressure measurement. 

e.Please keep the cuff clean. If the cuff becomes dirty, remove it from the monitor and clear it by hand 

in a mild detergent, then rinse it thoroughly in cold water. Never dry the cuff in clothes dryer or iron it. 

Clean the cuff after the usage of every 200 times is recommended. 

 

  TAKING YOUR BLOOD PRESSURE READING 

a.  Press the START button to start off-line measurement when disconnected with iOS device. If 

connected with iOS device.your monitor can be controlled by ios device 

b.  A beep is heard and all display characters are shown for self-test. See picture 6. Please contact 

the service center if segment is missing. 

       

       

 

                                       

 Picture 6            Picture 6-1         Picture 6-2 

c.  Then the current memory bank (U1 or U2) is blinking. See picture 6-1. Press “MEM” button to 

change over to other bank. See picture 6-2. Confirm your selection by pressing “START” button. 
The current bank can also be confirmed automatically after 5 seconds with no operation. 

d.  After selecting the memory bank, the monitor starts to seek zero pressure. See picture 6-3. 
 

       

       

 

     Picture 6-3

               

Picture 6-4          Picture 6-5 

e.  The monitor inflates the cuff until sufficient pressure has built up for a measurement. Then the 

monitor slowly releases air from the cuff and carries out the measurement. Finally the blood 
pressure and pulse rate will be calculated and displayed on the LCD screen separately. Irregular 
heartbeat symbol (if any) will blink. See picture 6-4&6-5. The result will be automatically stored in 
the current memory bank. 

f.  After measurement, the monitor will turn off automatically after 1 minute of no operation. 

Alternatively, you can press the “START” button to turn off the monitor manually or control your 
monitor to test again by your app. 

g.  During measurement, you can press the “START” button to turn off the monitor manually. 

 

Note: Please consult a health care professional for interpretation of pressure measurements. 

   

 

Summary of Contents for KD-557BT

Page 1: ...DING 3 MEASUREMENT PROCEDURES 4 TAKING YOUR BLOOD PRESSURE READING 5 DISPLAYING STORED RESULTS 6 DELETING MEASUREMENTS FROM THE MEMORY 6 SPECIFICATIONS 7 GENERAL SAFETY AND PRECAUTIONS 8 TROUBLESHOOTING 9 CARE AND MAINTENANCE 11 WARRANTY INFORMATION 11 EXPLANATION OF SYMBOLS 12 CONTACT AND CUSTOMER SERVICE 13 IMPORTANT INFORMATION REQUIRED BY THE FCC 13 OTHER STANDARDS AND COMPLIANCES 14 ELECTROMA...

Page 2: ...Cuff INTENDED USE The iHealth Wireless Blood Pressure Monitor Electronic Sphygmomanometer is intended for use in a professional setting or at home and is a non invasive blood pressure measurement system It is designed to measure the systolic and diastolic blood pressures and pulse rate of an adult individual by using a technique in which an inflatable cuff is wrapped around the upper arm The measu...

Page 3: ...se devices should be V6 0 or higher The iHealth Wireless Blood Pressure Monitor is also compatible with a number of Android devices the Android version should be V4 0 or higher and RAM should be 1 0G or more For a complete list of compatible devices please visit the support page on www ihealthlabs com Cuf LCD START MEM button Blood Pressure Level Classification Irregular Heartbeat Low Battery Dias...

Page 4: ...nth or more to avoid relevant damage of battery leakage Avoid the battery fluid to get in your eyes If it should get in your eyes immediately rinse with plenty of clean water and contact a physician The monitor the batteries and the cuff must be disposed of according to local regulations at the end of their usage Connect to iOS Device via Bluetooth Apply the cuff and press the START button to test...

Page 5: ...with your feet flat on the floor without crossing your legs b Place your hand palm side up in front of you on a flat surface such as a desk or a table c The center of the cuff should be at your heart level Lying Down During Measurement a Lie on your back b Place your arm straight along your side with your hand palm side up c The cuff should be level with your heart Note Blood pressure can be affec...

Page 6: ...current memory bank U1 or U2 is blinking See picture 6 1 Press MEM button to change over to other bank See picture 6 2 Confirm your selection by pressing START button The current bank can also be confirmed automatically after 5 seconds with no operation d After selecting the memory bank the monitor starts to seek zero pressure See picture 6 3 Picture 6 3 Picture 6 4 Picture 6 5 e The monitor infla...

Page 7: ...o operation You can also press the button START to turn off the monitor manually DELETING MEASUREMENTS FROM THE MEMORY When any result is displaying keeping on pressing button MEM for three seconds all results in the current memory bank will be deleted after three beep LCD will show picture 8 Press the button MEM or START the monitor will turn off Picture 8 Remember to 1 Make sure that the appropr...

Page 8: ...essure Monitor 2 Model KD 557BT 3 Classification Internally powered Type BF applied part IPX0 No AP or APG Continuous operation 4 Machine size 136 mm 100 mm 65mm 5 11 32 x 3 15 16 x 2 9 16 5 Cuff circumference 22cm 30cm 8 21 32 11 13 16 6 Weight about 236g 8 5 16 oz exclude batteries and cuff 7 Measuring method oscillometric method automatic air inflation and measurement 8 Memory volume 2x60 times...

Page 9: ...matically It is thus recommended that the phone be set in Airplane mode during measurement to avoid interrupting the measurement 6 If an Irregular Heartbeat IHB is detected during the measurement procedure the IHB symbol will be displayed Under this condition the Wireless Blood Pressure Monitor can keep functioning but the results may be inaccurate Please consult your physician for accurate assess...

Page 10: ...or APP display reads ERROR Fail to detect diastolic pressure LCD shows Er 3 or APP display reads ERROR Pneumatic system blocked or cuff is too tight during inflation Apply the cuff correctly and try again LCD shows Er 4 or APP display reads ERROR Pneumatic system leakage or cuff is too loose during inflation LCD shows Er 5 or APP display reads ERROR Cuff pressure above 300mmHg Measure again after ...

Page 11: ...第 10 页 共 16 页 ANDPRECAUTIONS No response Incorrect operation or strong electromagnetic interference Take out batteries for five minutes and then reinstall all batteries ...

Page 12: ...nded that if the cuff is used in a hospital or a clinic it be disinfected twice a week Wipe the inner side the side that contacts skin of the cuff with a soft cloth lightly moistened with Ethyl alcohol 75 90 Then air dry the cuff Do not drop this monitor or subject it to strong impact Avoid high temperature and direct sunlight Do not immerse the monitor in water as this will result in damage to th...

Page 13: ...UMBER Symbol for EUROPEAN REPRESENTATIVE Symbol for COMPILES WITH MDD93 42 EEC REQUIREMENTS iHealth is a trademark of iHealth Lab Inc Made for iPod Made for iPhone and Made for iPad mean that an electronic accessory has been designed to connect specifically to iPod iPhone or iPad respectively and has been certified by the developer to meet Apple performance standards Apple is not responsible for t...

Page 14: ...e in a residential installation This product generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instructions may cause harmful interference to radio communications However there is no guarantee that interference will not occur in a particular installation If this product does cause harmful interference to radio or television reception which c...

Page 15: ...nvironment Emissions test Compliance Electromagnetic environment guidance RF emissions CISPR 11 Group 1 The KD 557BT uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions CISPR 11 Class B The KD 557BT is suitable for use in all establishments other than domestic and those direct...

Page 16: ...s the recommended separation distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in ...

Page 17: ...e maximum output power of the communications equipment Rated maximum output power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d i...

Reviews: