MULTIGASSENS-М2 USER MANUAL
Rev. 1.8 07.02.19
19
Table 8
Interval between
data recording
Log file, seconds
Log time length at
continuous device
operation,
hours
Log duration during
continuous operation
of the device,
days
The duration of the log from
the device in shifts, the length
of the shift is 8 hours,
changes
15
49
2
6
30
99
4
12
60
198
8
24
240
792
33
99
3600
11891
495
1486
2.6. Setting device parameters
Adjustment of the parameters of the device is possible using an external program installed on a
device that supports data transmission via ISO 15693 (NFC), this device can be a smartphone with NFC
module, or a computer with an NFC data reader.
3
Maintenance and verification
3.1
Maintenance service
All maintenance operations of the device must be performed in compliance with the requirements for
protection from static electricity.
Maintenance of the device is divided into external inspection and preventive maintenance. During
the external examination it is necessary to perform a visual inspection in order to identify malfunctions
and malfunctions in the operation of the device.
At prophylactic maintenance of the device the following types of works are performed:
-
cleaning of dirt and dust buttons, protective filters, display, labels on nameplates;
-
check of structural elements for mechanical damage;
-
check for the seal.
Prophylactic maintenance is carried out every 3 months
3.2
Verification
Verification is made according to the document MP 98-221-2017 "Portable gas analyzers
MULTIGASSENS-M2. Calibration procedure ".
Calibration interval - 1 year.