STERYLIS ULTRA Operating Manual
CAUTION!
CAUTION!
CAUTION!
The replacement of UV-C fluorescent tubes in STERYLIS sterilisers may only be
carried out by the manufacturer’s service or another authorised service centre!
The replacement of ozone generators in STERYLIS sterilisers may only be
carried out by the manufacturer’s service or another authorised service centre.
The replacement of the ozone sensor in STERYLIS sterilisers may only be carried
out by the manufacturer’s service or another authorised service centre.
Replacing the ozone generators
Replacing the ozone sensor
Regular replacement of the UV-C lamps is essential for proper operation. The lifetime of the generators is
approx. 9000 hours.
The ozone sensor is one of the most important elements of the device, which determines its proper
functioning and safety. Regular replacement of the ozone sensors is essential for proper operation.
The expiry date of the ozone sensor is 12 months – after that time it must be replaced regardless of
the operating time of the device. The ozone sensor also has a service life of 4,500 hours. If the sensor’s
operating time is exceeded before 12 months – it must be replaced after 4,500 operating hours.