6.2 Installation instructions for operation without overflow
prevention
[MEdI] = [Auto] or [OP] = [OFF] (overflow prevention OP deactivated)
6.2.1 Installation in the inactive zone
Between the maximum level (b1) and the inactive zone (I1), minimum
distance (a1) must be adhered to (see Fig. 6-2 and Table 6-3).
►
Fix the unit using mounting elements
in the inactive zone (I1). The outside
length (c) must not exceed (I1) (Table
6-3).
►
Ensure that the maximum level (b1)
is not exceeded after completed
installation (Table 6-3).
►
Observe further minimum distances
according to Table 6-4.
Fig. 6-2
A
I1
c
I2
b1
a1
x
u
I1 / I2:
Inactive zones
A:
Active zone
a1:
Minimum distance between the
inactive zone (I1) and the maximum
level (b1)
b1:
Max. level from the lower edge of the
sensor (without offset)
c:
outside length
Table
6-3
LT8022
LT8023
LT8024
[cm]
[inch]
[cm]
[inch]
[cm]
[inch]
I1
5.3
2.1
6.0
2.4
10.4
4.1
A
19.5
7.7
39.0
15.4
58.5
23.0
a1
1.0
0.4
1.5
0.6
2.5
1
b1
20.0
7.9
39.5
15.6
59.5
23.4