12 Operation
After switch-on of the operating voltage, the unit is in the operating mode (= nor-
mal operating mode). It carries out its measurement and evaluation functions and
generates output signals according to the set parameters.
►
Check whether the unit operates correctly.
12.1 Operation indication
[----] (continuous)
Initialisation phase after power on
[numerical value] +
LED 1
Current level in cm
[numerical value] +
LED 2
Current level in inches
LED 7 / LED 8
Switching status OUT2 / OUT1 (on if output x is switched)
[----]
Level below the active zone
[FULL] + [numerical
value] alternately
The overflow prevention OP is reached (overflow warning) or
the level is above the active zone.
Adjustment [cOP] of the overflow prevention OP necessary.
[Loc]
Unit locked via buttons; parameter setting is not possible. For
unlocking press the two setting buttons for 10 s.
[uLoc]
Unit is unlocked / parameter setting is possible again.
[C.Loc]
The unit is temporarily locked. Parameter setting via IO-Link is
active (temporary locking).
[S.Loc]
Unit is permanently locked via software. This locking can only
be removed with a parameter setting software.
12.2 Read set parameters
►
Briefly press [Mode/Enter] (if required, repeat several times).
> Menu structure is scrolled until the required parameter has been reached.
►
Briefly press [Set].
> Respective parameter value is displayed for 30 s.