7.6.1 Installation to G¾ process connection directly in the tank lid
The elastomer seal on the sensor is used as
process seal. The supplied flat seal can be used
to smooth out unevenness on the tank lid process
connection.
The upper sealing area on the process connection
must be flush with the tapped hole.
►
Lightly grease the sensor thread with a suitable
paste.
►
Insert the unit into the process connection.
► Tighten it using a spanner. Tightening torque:
35 Nm.
Fig. 7-10
R
7.6.2 Installation in the tank lid using a G¾ flange plate
Fig. 7-11
Fig. 7-12
d
A
R
►
Arrange for a bore hole in the tank lid. It must have a minimum diameter d to
enable sufficient transfer of the measured signal to the probe (Fig. 7-11). The
diameter (d) depends on the wall thickness of the tank lid:
Wall thickness [mm]
1...5
5...8
8...11
d [mm]
35
45
55
► Install the flange plate with G¾ process connection (→ Accessories) with the
flat surface showing to the tank and fix it with appropriate screws.
If necessary, a seal (A in Fig. 7-12) can be inserted between flange plate
and tank. Some flange plates are supplied with a seal. If this is not the
case, use a suitable seal.