22
11.4 Fault indications
[SC1]
Short circuit in OUT1�
[SC2]
Short circuit in OUT2�
[SC]
Short circuit in both outputs�
[OL]
Detection zone of volumetric flow or temperature exceeded�
Measured value between 120 % and 130 % of the final value of the measuring
range�
[UL]
Below the temperature detection zone: measured value below -10°C (14 °F)�
[Err]
•
Unit faulty / malfunction�
•
Measured value greater than 130 % of the final value of the measuring range�
[SEnS] Sensor indicates incorrect measurement� Possible cause: accumulated gas and
air in the medium or unit�
For more detailed diagnostics / fault assessment:
►
Briefly press [Set]�
> The latest measured values are displayed�
[IOE]
Flow sensor faulty
[Loc]
Setting pushbuttons locked, parameter change rejected�
12 Technical data
Technical data and scale drawing at www�ifm�com�