PRS-300 Series
ETHERNET INTERFACE OPTION
18
Figure 4-4 9006 Con
fi
guration Page
7. Enter the new settings as desired. If you select
DHCP for the TCP/IP Mode, the page blanks out
the IP, Net and Gateway addresses as they will
be supplied by your DHCP server. Check the
entered values carefully as the unit’s webserver
does minimal error checking. Press the ‘Update
Flash’ button when done. A Con
fi
rmation Page
similar to the one shown in Figure 4-5 will appear
in your browser.
Figure 4-5 9006 Con
fi
rmation Page
8. Your new settings have been saved in the
board’s
fl
ash memory. You have to reboot the unit
or power cycle it for the changes to take affect.
Press the ‘Reboot’ button to reboot the unit now
or the ‘Return to the Con
fi
guration Page’ button
to revisit the new settings.
5.6 VXI-11 Configuration Utility
Method
The VXI-11 Configuration Utility program
is called ‘VXI11_con
fi
g.exe’ and runs in any
WIN32 PC with Windows 98, Me, 2K, XP,
Vista, 7 and Server 2003 operating systems. The
VXI11_con
fi
g.exe program can be found on the
PRS product page at www.ietlabs.com.
The VXI11_con
fi
g program can be run from the
CD or can be installed onto your hard disk and
run from the installation directory.
VXI11_con
fi
g.exe is a Visual Basic program and
requires that either Microsoft’s Visual Studio 6
or VBruntime6 be installed on your PC to run.
1. Connect the board directly to the WIN32 PC
that will be running the Con
fi
guration Utility.
Disconnect the PC from the company network
and use the supplied Ethernet Crossover Cable to
connect the PC to the board as shown in Figure
4-6. This will eliminate any potential network
con
fl
icts while con
fi
guring the board.
g
g
Figure 4-6 9006 Connected to PC with a
Crossover Cable
Alternately, use a standard Ethernet patch cable
Summary of Contents for PRS-300 Series
Page 2: ...TEL 516 334 5959 800 899 8438 FAX 516 334 5988 www ietlabs com IET LABS INC...
Page 16: ...SPECIFICATIONS PRS 300 Series 6...
Page 24: ...PRS 300 Series IEEE INTERFACE OPTION 14...
Page 34: ...PROGRAMMING PRS 300 Series 24...
Page 42: ...PRS 300 Series PRS DMM SOFTWARE 32...
Page 50: ...PRS 300 Series APPENDIX B 40...