VI-700 V&l Source
3
SPECIFICATIONS
Figure 2.1: Typical OPERATING GUIDE Affixed to Unit
2.5
HI - High Current Option - 200 mA
Output
Output Current:
200 mA.
Resolution:
0. 1 mA.
Accuracy:
±(0.01%+0.75 LSD)
Compliance:
0- 18 V; with LED overload indicator.
2.6 REMOTE Features:
Front panel switch selects
REMOTE
(digital in-
terface) or
LOCAL
(front panel thumbwheel
switch) operation. All front panel controls except
POWER
are programmable.
IEEE-488.2 Interface Option:
IEEE-488.2-1987; SCPI 1994.0;
"*
IDN"
for S/N, Model & REV;
"
CAL:DATe?
" for last calibration date.
RS232 Interface Option:
RS232.SCPI 1994.0;
"*
IDN"
for S/N, Model & REV;
"
CAL:DATe?
" for last calibration date;
25 pin male interface conforming to EIA-STD-
RS-530.
CURRENT
OUTPUT RANGES
RESOLUTION
ACCURACY
COMPLIANCE
OUTPUT RANGES
RESOLUTION
ACCURACY
OUTPUT CURRENT
OUTPUT IMPEDANCE
NOISE and RIPPLE
TEMPCO
CONSULT INSTRUCTION MANUAL FOR PROPER INSTRUMENT OPERATION
VI-700 OPERATING GUIDE
VOLTAGE
1.
CNTRL MODE
button has no effect for non programmable
units.
2. Make the most secure connection to the V terminals if
significant current is drawn.
3. Current
OVLD
LED indicator is on if
mA
terminals are open.
This is of no concern
or leads may be shorted together.
4. The ground terminal is connected to earth ground and may
be connected to either
VOLTAGE
terminal.
MODEL: VI-700 SN: J1-0540317
OPERATION
POWER
FUSE
105-125 V, 50-60 Hz; 5 W.
0.25 A
200 mV, 2 V, 20 V
100 µV, 1 mV, 10 mV
±(75 ppm + 50 µV); 200 mV range;
±(75 ppm + 0.25 mV), 2 V range;
±(75 ppm + 2.5 mV), 20 V range;
at 23°C, no load, after warmup.
70 mA, maximum load; with LED
overload indicator
5 m
Ω
typical, 40 m
Ω
for 20 V range
50 µVrms on 200 mV and 2 V
ranges, 100 µVrms on 20 V range
(exclusive of random transients)
±10ppm/°C typical, ±15ppm/°C max.
VILBL/p2/VI-700Rev J3/90-100%12-04
Traceable to SI
IET LABS, INC.
534 Main Street, Westbury, NY 11590 • (800) 899-8438 • (516) 334-5959
CAGE CODE: 62015
www.ietlabs.com
200 µA, 2 mA, 20 mA
0.1 µA, 1 µA, 10 µA
±(75 ppm + .05 µA); 200 µA;
±(75 ppm + 0.25 µA), 2 mA;
±(75 ppm + 2.5 µA), 20 mA;
at 23°C, no load,
after warmup.
0-20 V; with LED overload
indicator
Summary of Contents for VI-700