PRS-DMM MODE
PRS-370 Series
25
Figure 4-5 Address
Use the key to move UP and the to select the
GPIB address that the Fluke 8508A or Keysight
3458A DMM is set to and then press
ENTER.
4.3 Performing an Adjustment
Once all of the PRS-DMM setting have been set
press the BACK button to go to the Home Screen.
The top left of the screen indicates PRS-DMM,
Resistance mode. The bottom half of the screen
shows the DMM being used and address. See
Figure 4-6.
Figure 4-6 PRS-DMM Home Screen
Touching the resistance value brings up the entry
screen for resistance or the numeric keypad can
be used to enter the required resistance value as
shown in Figure 4-7.
Figure 4-7 PRS-DMM Enter resistance value
screen
Enter the required resistance value and press one
of the unit buttons.
To initiate an adjustment make sure IEEE and
measurement cables are connected between the
DMM and the PRS-370. Make sure the DMM is
turned on and then press the
DMM
button shown
in Figure 4-6.
The Adjusting Screen will be shown in
fi
gure 4-8
during the adjustment process.
Figure 4-8 Adjusting Screen
Once the adjustment process is complete the
screen shown in Figure 4-9 will be shown.