background image

7

MAINTENANCE

HARS  Series

Chapter 4

MAINTENANCE

4.1

Verification of Performance

4.1.1 Calibration Interval

The 

HARS 

Series instruments should be verified

for performance at a calibration interval of twelve
(12) months. This procedure may be carried out by
the user if a calibration capability is available, by IET
Labs, or by a certified calibration laboratory.

If the user should choose to perform this procedure,
then the considerations below should be observed.

4.1.2 General Considerations

It is important, whenever testing the 

HARS-X

 Se-

ries Decade Units, to be very aware of the capabili-
ties and limitations of the test instruments used. A
resistance bridge may be employed, and there are
direct-reading resistance meters or digital multim-
eters available that can verify the accuracy of these
units, especially when used in conjunction with stan-
dards that can serve to confirm or improve the ac-
curacy of the testing instrument

Such test instruments must be significantly more ac-
curate than ±(2 m

) for all applicable

ranges, allowing for a band of uncertainty of the in-
strument itself. A number of commercial bridges and
meters exist that can perform this task; consult IET
Labs.

It is important to allow both the testing instrument
and the 

HARS-X

 Substituter to stabilize for a num-

ber of hours at the nominal operating temperature of
23

O

C, and at nominal laboratory conditions of hu-

midity. There should be no temperature gradients
across the unit under test.

Substantial Kelvin type 4-wire test terminals should
be used to obtain accurate low-resistance readings.
It is convenient, once the zero resistance has been
determined, to subtract it from the remaining mea-
surements. This can be automatically done in many
instruments which have an offset subtraction capa-
bility.

4.1.3 Procedure

1. Confirm the zero resistance of the unit.
2. Determine the allowable upper and lower

limits for each resistance setting of each decade
following the specified accuracy.  For the

HARS-X 

series, these limits for any resistance

“R” are [R±(0.0001 R + 2 m

)]. For the A,

B, or Q series,  see  specifications.

3. Confirm that the resistances fall within these

limits after subtraction of the zero resistance.

4. If any resistances fall outside these limits, the

associated switch assembly may require
replacement.

4.2  Schematic

Refer to Figure 4.1 for a schematic of the 

HARS

decade unit.

Summary of Contents for HARS-X Series

Page 1: ...PRECISION INSTRUMENTS FOR TEST AND MEASUREMENT Copyright 2002 IET Labs Inc HARSim January 2002 TEL 516 334 5959 800 899 8438 FAX 516 334 5988 www ietlabs com IET LABS INC in the GenRad 534 Main Street...

Page 2: ......

Page 3: ...eplaced at no charge when returned to IET Changes in this product not approved by IET or application of voltages or currents greater than those allowed by the specifications shall void this warranty I...

Page 4: ...Wire Kelvin Lead Connections 5 3 2 4 Thermal emf Considerations 6 3 3 Dial Setting 6 3 4 Environmental Conditions 6 Chapter 4 MAINTENANCE 7 4 1Verification of Performance 7 4 1 1 Calibration Interval...

Page 5: ...WHENEVER HAZARDOUS VOLTAGES 45 V ARE USED TAKE ALL MEASURES TO AVOID ACCIDENTAL CONTACT WITH ANY LIVE COMPONENTS USE MAXIMUM INSULATION AND MINIMIZE THE USE OF BARE CONDUCTORS REMOVE POWER WHEN HANDL...

Page 6: ......

Page 7: ...nstruments Contact resistance is stable and remains low and repeatable High qualitygold platedtellurium copperbindingposts serve to minimize the thermal emf effects which would artificially reflect a...

Page 8: ...tance Total Stability Long Term Temperature Max Maximum Maximum per Decade Stability Coefficient Power current voltage Step Resistance ppm year ppm 3 years ppm C W step per decade per step 1 m 10 m 10...

Page 9: ...ircuits or other con figurations Each unit consists of low inductance resistors in series with a high perfor mance solid silver alloy contact switch For less exacting applications more economical tole...

Page 10: ...istance at 23 C traceable to NIST Zero Resistance 1 m per decade typical at dc Power per Step 25 mW for 1 m steps 0 2 W for 10 m steps 0 25 W for 100 m steps 0 6 W for 1 through 100 steps 0 5 W for 1...

Page 11: ...thenecks of the binding posts If the four terminals are avail able separately the optimal connection is shown in Figure 3 1 where the current leads are introduced into the top of the binding posts and...

Page 12: ...position on any one decade equals the 1 position on the next higher decade It adds about 11 to the nominal total decade resistance To determine the resistance obtained when one or more 10 settings are...

Page 13: ...funcertaintyofthein strumentitself Anumberofcommercialbridgesand meters exist that can perform this task consult IET Labs It is important to allow both the testing instrument andtheHARS XSubstituterto...

Page 14: ...8 MAINTENANCE HARS Series Figure 4 1 HARS Series Schematic Diagram...

Page 15: ......

Reviews: