Safety
12
3.3
Important device instructions
ATTENTION
Equipment failure
▪
The device must not be used in the vicinity of
magnetic resonance imaging apparatus or in
the direct proximity of another electrical
medical monitor.
▪
The device is not suitable for simultaneous
use with high frequency surgery monitors.
▪
Do not drop the blood pressure monitor and
do not place objects on top of it.
▪
Do not use the device directly adjacent to
other devices or stacked with other devices,
as this may result in malfunction. If operation
in the manner described above becomes
necessary nevertheless, this device and the
other devices should be monitored to ensure
that they are functioning correctly
▪
The use of components not included in the
scope of delivery may lead to measurement
errors. Thus, only use the accessories offered
by I.E.M. GmbH.
▪
The cuff and the hose are made of a material
that does not conduct electricity. They thus
protect the device against the effects of
discharging a defibrillator. In the event of
discharging a defibrillator, the device itself
must not touch the patient since the device
can be damaged as a result of such a
discharging and can result in the incorrect
value being displayed.
ATTENTION
Warranty
▪
Do not open the housing of the Tel-O-Graphs
®
BT, otherwise any warranty becomes void.