Troubleshooting
103
EN
Note
Some errors are combined with a continuous alarm for safety reasons. The continuous alarm can be ended
by pressing any button. If there is residual pressure inside the cuff, open the cuff immediately.
5.4 Mobil-O-Graph
®
error description
Error message
Possible cause
Measures
Err 1
1. The patient has severe
arrhythmia
1. Blood pressure monitor not
applicable
2. Arm moved during
measurement
2. Keep the arm still during
measurement
3. Insufficient valid pulse rate
detected
3. Place the cuff on the arm again
Err 2
1. Arm moved during
measurement
1. Keep the arm still during
measurement
2. Cuff does not fit the arm snugly
2. Check the seating of the cuff and
that of the device