P9242-R-EVK Evaluation Board User Manual
© 2016 Integrated Device Technology, Inc.
4
December 22, 2016
1.
Setup
1.1
Required or Recommended User Equipment
The following additional lab equipment is required for using the kit:
P9221-R-EVK Receiver Evaluation Board or any WPC-1.2.2 compliant receiver
Power supply or 12V/2A AC adaptor
1.2
Kit Hardware Connections
Follow these procedures to set up the kit as shown in Figure 1.
1.
Set up the P9242-R Evaluation Board: Plug the 12V adapter or user’s power supply into the J3 (barrel connector on the bottom of the
board) or J1 (micro USB connector).
2.
Place the P9221-R-EVK or the user’s receiver on the transmitter (TX) pad with the components facing up as shown in Figure 1.
3.
Verify that the two green LEDs identified in Figure 1 are illuminated indicating that coupling has been established.
4.
If using the P9221-R-EVK Evaluation Board as the receiver, connect wires to the VOUT and GND test points on the P9221-R-EVK
receiver to measure the output voltage and apply a load.
Figure 1. Evaluation Kit Connections using the P9221-R Receiver Evaluation Board
P9242-R Transmitter Evaluation Board
Transmitter Coil
P9221-R-EVK
Green LEDs illuminate when connection has been established.