5-5-1- Decontaminating the containers of solid and liquid waste
WARNING:
RISK OF BIOLOGICAL CONTAMINATION
This part of the analyser is in contact with biological samples. There is, therefore, a
potential risk of infection.
Wear disposable gloves for all handling procedures.
Solid waste tray
•
Empty and decontaminate the waste tray once emptied with a disinfectant solution suitable for medical
devices.
•
Rinse it with water, then decontaminate again with a disinfectant solution suitable for medical devices.
•
Dry the tray prior to replacing in its compartment.
Liquid waste container
•
Rinse the container with freshly prepared bleach once the liquid waste has been disposed of.
•
Then rinse with tap water.
•
Add the appropriate volume of bleach prior to reusing the container with the analyser (see Section
Appendices,
A-1- Waste Disposal,
page A2).
Maintenance
5-5-
Analyser Cleaning
(continued)
IDS-iSYS User Manual - Revision M1
Maintenance 5-5
147
Software version V 14