SPECIFICATIONS
Mode of Operation:
For continuous operation with short time loading
Duty Cycle: 10% (6 min/hr)
Type of Equipment:
Class I Type B (as defined by IEC 60601-1, UL 60601-1,
EN 60601-1, CSA 601.1, IEC 60601-2-32:1994 and IEC 60601-2-46)
Type B protection against electrical shock
Complies with FDA standards
Electrical:
Supply Voltage: 120±5% Vac 60Hz or 230
10% Vac 50Hz
Duty Cycle: 10% (6 min/hr)
Current Rating: Less than 10 Amps
Internally Battery Powered
Environmental:
Operating Temperature Range: -10
C to +40
C
Operating Humidity Range: 30% to 75% relative humidity, noncondensing
Operating Pressure Range: 700hPa to 1060 hPa
Transport & Storage Temperature Range: -40
C to +60
C
Transport & Storage Humidity Range: 30% to 75% relative humidity, noncondensing
Transport & Storage Pressure Range: 500hPa to 1060 hPa
Rated IPX4 (Protected against splashing water)
Meets EMC requirements of EN 60601-2 (1993)
Tabletop:
The tabletop is made of carbon fiber and is certified to meet all the requirements of 21
CFR Subchapter J