IMAGING END
OF TABLETOP
Aspect 100-4T Q-CARD
SETUP/PATIENT PREPARATION:
Moving table (NEVER WITH PATIENT ON BOARD):
1) Unlock Rear Casters using Caster Lock Pedal.
2) Unlock Front Steering using Front Steer Lock Pedal.
3) Position table to desired location.
4) Secure table mobility by locking Casters and front steering
using Rear Caster Lock Pedal and Front Steer Lock Pedal.
AC Power:
5) Plug power cord into properly grounded outlet.
6) Protect power cord from damage by casters or other
equipment.
Controls:
1) Emergency Stop Pushbutton disables all tabletop motion
control when engaged.
2) If tabletop motion controls fail to operate, make sure
that the Emergency Stop Pushbutton is NOT pushed in.
BATTERY OPERATION (DC POWER):
1) The table is FULLY functional, but motions are slower.
2) Connect to AC power for a minimum of 8 hours daily to
ensure a proper battery charge.
3) Failure to push in Emergency Stop Pushbutton when table
is not in use and not connected to AC power for longer
than 72 hours, may result in permanent damage to the
batteries.
CPR:
Return table to level, retract tabletop to minimize
overhang and lower table position to a comfortable
height to perform CPR.
CLEANING AND STORAGE:
1) Unplug power cord and wrap it around
the cord holder.
2) Push in Emergency Stop Pushbutton.
3) Lock rear casters and front steering.
4) Clean and disinfect table.
WARNING! Read operator
manual prior to use
CONNECTION PORT FOR HAND
OR FOOT CONTROLS CONNECT
HERE
EMERGENCY
STOP
PUSHBUTTON
REAR CASTER LOCK PEDAL
BRAKES:
The 100-4T table braking system utilizes a passive magnetic brake;
it remains engaged in the event of a complete loss of electrical
power. Power is necessary to release brakes.
Braking force is limited by the magnet strength. Use caution when
applying pressure to the side of the table as the brakes may be
overcome if more than 50 pounds of force is applied.
To reduce potential motion during patient transfers with larger
patients, release the table brake and move the top in the direction
the patient will be pulled during loading or unloading until the
tabletop transverse travel limit is reached. Release the panning
handle to re-engage the lock and then transfer the patient.
POWER CORD
HOLDER
PEDESTAL END OF TABLETOP