Aspect ISR G3 Q-CARD 

SETUP/PATIENT PREPARATION: 

Moving table: 

  To move table, set caster lock pedal to FLOAT or 

STEER position. 

  To secure table mobility, set caster lock pedal to LOCK 

position. 

AC Power: 

1)  Plug power cord into properly grounded outlet. 
2)  Protect cord from damage by casters or other 

equipment. 

Controls: 

1)  Emergency Stop Pushbutton disables all tabletop motion 

control when engaged. 

2)  If tabletop motion controls fail to operate, make sure 

that the Emergency Stop Pushbutton is not pushed in. 
 

BATTERY OPERATION (DC POWER): 

1)  The table is FULLY functional, but motions are slower. 
2)  Connect to AC power for a minimum of 8 hours daily to 

 ensure a proper battery charge. 

3)  Failure to push in Emergency Stop Pushbutton when 

table is not in use and not connected to AC power for 
longer than 72 hours, may result in permanent damage 
to the batteries. 
 

CPR: 

Return table to level, retract tabletop to minimize 
overhang and lower table position to a 
comfortable height to perform CPR. 

 

  CLEANING AND STORAGE: 

1)  Unplug power cord and wrap it around cord holder. 
2)  Push in Emergency Stop Pushbutton. 
3)  Set caster lock pedal to LOCK position. 
4)  Clean and disinfect table. 
5)   

WARNING! Read operator 

manual prior to use 

CONNECTION POINT FOR 
HAND OR FOOT CONTROLS 
CONNECT HERE 

EMERGENCY 

STOP 

PUSHBUTTON 

CASTER LOCK PEDAL 

www.imagediagnostics.com 

BRAKES: 

The ISR G3 table braking system utilizes a passive magnetic brake; 
it remains engaged in the event of a complete loss of electrical 
power. Power is necessary to release brakes.

 

Braking force is limited by the magnet strength. Use caution when 
applying pressure to the side of the table as the brakes may be 
overcome if more than 70 pounds of force is applied. 

To reduce potential motion during patient transfers with larger 
patients, release the table brake and move the top in the direction 
the patient will be pulled during loading or unloading until the table 
top transverse travel limit is reached. Release the panning handle to 
re-engage the lock and then transfer the paitent. 

POWER 

CORD 

RECEPTACLE 

PEDESTAL END OF TABLETOP 

IMAGING END OF TABLETOP 

L100-2859A        03/2018 