It is recommended that the pendant
control be disconnected during critical
procedures to prevent inadvertent
operation or motion, causing patient
injury. The Motion Lockout Pushbutton
on the Panhandle Unit does
not
disable
the Handset Control.
CASTER LOCK
PEDAL POSITIONS
FLOAT
(Movable in
all directions)
STEER
(Rear wheels
roll forward and
backward only)
LOCK
(Will not move in
any direction)
•
PUSH IN to stop
ALL
tabletop motion.
•
TWIST to reset.
ELEVATE
SWITCH
TRENDELENBURG
SWITCH
(Green light to left of
switch indicates table
top is level for this
motion)
SWITCH
LOCKOUT
PUSHBUTTON
(Amber light
indicates that
switches on pan
handle are
disabled)
PANHANDLE
KNOB/ TABLETOP
FLOAT BRAKE
(Lift underside of
Panhandle Knob to
float table top. It will
not operate if table
top is not level)
L100-2912 Rev.B
Table Top Motion
Handset Control
Patients must be loaded from the side of table. Tilting or instability may occur if
patient is loaded onto the pedestal end of table or the imaging end of the table.
EMERGENCY
STOP
PUSHBUTTON
(Green light to
right of Tabletop
Float Brake
indicates table top
is centered on the
base laterally)
TABLE TOP MOTION
CONTROL PANHANDLE
Elevate
Tilt up Imaging
End
(Trendelenburg)
Trendelenburg
Level Indicator
Laterally Centered
Indicator
Lower
Tilt Down
Imaging End
(Trendelenburg)
Aspect 100-4T G3 Q-CARD