Operator Manual
Aspect 100-4
L100-3316 Rev. A
12
•
When the table is not connected to AC power, the table will automatically switch to backup
battery mode. The table should only be used under backup battery power temporarily for
necessary function if the external power supply is lost.
•
It is recommended that AC power be applied for a minimum of 8 hours every day to keep a
proper charge on the batteries and achieve maximum battery life. Batteries are constantly
being charged during normal use when connected to AC power.
When the table is not in use and not connected to AC power, the Emergency
Stop Pushbutton must be fully engaged to prevent the batteries from being
discharged and to safely power off the equipment.
•
The
Aspect
100-4 tabletop braking systems utilize a passive magnetic brake; it remains
engaged/energized in the event of a complete loss of electrical power. Power is necessary to
release the brakes.
•
As the braking system utilizes permanent magnetic brakes, braking force is limited by the
magnet strength. Use caution when applying pressure to the side of the table as the brakes can
be overcome if more than 70 pounds of force is applied.
7.2.
Preparation for Patient Use
•
New Installation: This equipment will need to be properly cleaned before patient use as it will
inevitably come into contact with contaminants during shipping, unpacking, storage, and
installation.
•
After Installation: This equipment will need to be properly cleaned between uses with patients
as it will inevitably come into contact with contaminants during procedures. Refer to section
11 of this manual for cleaning instructions and approved cleaning substances.
7.3.
Patient Loading
CAUTION
If the tabletop is struck hard enough, it may slide
to reduce damage to the carbon fiber top.