Appendix B: Air supply requirements
Compressed air is required for the Filta-Max
xpress
Pressure Elution Station to function properly
because it supplies the air pressure needed to elute the Filter Modules. IDEXX supplies a hose kit
that lets you connect your facility’s air supply to the Pressure Elution Station. The hose kit consists
of a 12-foot self-retracting hose and connector.
Note:
The air supply should be set to a minimum pressure of 5.0 bar (72.5 psig) and a maximum of
8.0 bar (116 psig).
For details regarding the air supply, refer to pages 10–11 or call IDEXX Technical Service.
Outlet air line with
supply valve
Pressure regulator
adjustment tap
Receiver air
pressure gauge
Outlet air
preparation filter
Outlet air
pressure gauge
Receiver drain tap
Air compressor unit
Regulator
Compressor air
inlet filter
Receiver (15 L)
Pressure
switch
Power
cord
On/off
switch