BioTek* 405*R 
 

IDEXX Technical Service washer maintenance 

 

Important points 

1.

 

There are two evening maintenance protocols. The protocol you use depends on how frequently you use the 
washer:   

 

If the instrument is in constant use, run the protocol that keeps the tubes primed and wet when the instrument is 

not being used.  

 

If the instrument is used only once or twice a week, run the protocol that purges the water. 

2.

 

Do not leave buffer in the system for long periods of time (for example, overnight or over the weekend). Doing so will 
clog the pins, and you will need to remove the manifold and clean it with a stylus. 

3.

 

If you must clean the manifold with a stylus, be sure to flush the manifold with running deionized (DI) water through 

the large port holes to remove any debris. If this step is not performed, the particles will immediately clog the  
pins again. 

Always rinse at the beginning of the day 

1.

 

Connect hose to DI water.   

2.

 

Turn power on. 

3.

 

Press 

Main Menu

.  

4.

 

Select 

MAINT 

and then 

DAY_RINSE

 (rinse will be with DI water). 

5.

 

Press 

ENTER

.  

6.

 

Press 

START

7.

 

Repeat DAY_RINSE (steps 3–6).

 

 

8.

 

Reconnect hose to the appropriate wash buffer for testing. 

9.

 

Press 

Main Menu

 

10.

 

Select 

MAINT 

and then 

DAY_RINSE

 (rinse will be with the wash buffer) 

11.

 

Press 

ENTER

12.

 

Press 

START

Evening maintenance, if washer will be used the next day 

1.

 

Connect hose to DI water. 

2.

 

Press 

Main Menu

.  

3.

 

Select 

MAINT 

and then 

DAY_RINSE

 (rinse will be with DI water). 

4.

 

Press 

ENTER

.  

5.

 

Press 

START.

 

6.

 

Select 

Main Menu

7.

 

Select 

MAINT

8.

 

Use the Option button to find the 

RINSE_AND_SOAK

 protocol. 

9.

 

When the wash manifolds are submerged, turn off instrument. 

 

 

 