5.2.
Measurement Options
This menu enables the selection of options for temperature measurements. The symbols
shown below indicate the icons shown for Function Key 1 when the various options are
selected.
♦
Cursors
.
Choose between one or two cursors. When two cursors are chosen, the
temperatures at both cursors and the temperature difference between them will be
displayed. Use Function Key 1 to select one of the cursors, which can then be moved
around on the display by the navigation buttons.
♦
Tracking
. Select “High”, “Low”, or “High & Low” in order to track and measure the hottest
point, the coldest point, or both hottest and coldest points in the image.
♦
Area
. If this option is turned on, the highest, lowest, and average temperatures within
the designated area will be displayed. Three different area size boxes can be selected
via Function Key 1
Note that the above three items are mutually exclusive, i.e. when one is turned on the
other two are disabled.
♦
Isotherms
. Select “High”, “Low”, or “High & Low” in order to highlight areas of the
scene with temperatures within one or two temperature bands. The temperature bands
are adjustable by means of Function Key 1 and the navigation keys.
♦
Temp. Profile
. Select “Horizontal” or “Vertical” to enable a histogram of temperature values
along a horizontal or vertical cross section to be displayed on the right hand side of the
display. The position of the cross section is indicated by small arrows at the left and
right or top and bottom of the image and can be adjusted by means of Function Key 1
and the navigation buttons.
♦
Temp alarms