- 12 -
function
Calibrator
Output
Reading
Low limit
High limit
LPF ACV
2V ,50Hz
1.5V
2.5V
54V ,50Hz
52.7V
55.3V
66V ,50Hz
64.5V
67.5V
300V,50Hz 295.0V 305.0V
600V,50Hz 590.5V 609.5V
2V ,400Hz
1.4V
2.6V
54V ,400Hz
50.8V
57.2V
66V ,400Hz
62.2V
69.8V
300V,400Hz 284.5V 315.5V
600V,400Hz 569.5V 630.5V
Table 4-2 Testing the Earth-bond Function
function
Applied
Reading
Low limit
High limit
Ω
2.00
Ω
1.92
Ω
2.08
Ω
36.00
Ω
35.41
Ω
36.59
Ω
360.0
Ω
354.3
Ω
365.7
Ω
3600
Ω
3543
Ω
3657
Ω
36.00k
Ω
35.43k
Ω
36.57k
Ω