![IDEAL 61-737 Operation And Safety Manual Download Page 22](http://html1.mh-extra.com/html/ideal/61-737/61-737_operation-and-safety-manual_616662022.webp)
22
Electrical Specifications
Function
Range
Resolution
Accuracy
61-737
±(a%+b)
AC Voltage (V)
4.000V
0.001V
±(1.0%+5)
40.00V
0.01V
400.0V
0.1V
600V
1V
±(1.5%+5)
AC Current (A)
40.00A
0.01A
±(2.0%+9)
400.0A
0.1A
±(2.0%+5)
Frequency (ACA Hz)
99.99Hz
0.01Hz
±(0.1%+5)
999.9Hz
0.1Hz
DC Voltage (V)
400.0mV
0.1mV
±(1.0%+8)
4.000V
0.001V
±(1.0%+5)
4.000V
0.01V
400.0V
0.1V
600V
1V
±(1.5%+5)
Resistance (
Ω
)
400.0
Ω
0.1
Ω
±(1.5%+5)
4.000k
Ω
0.001k
Ω
40.00k
Ω
0.01k
Ω
400.0k
Ω
0.1k
Ω
4.000M
Ω
0.001M
Ω
40.00M
Ω
0.01M
Ω
±(2.0%+5)
1. Overload Protection: 600VRMS and 400 AAC
2. Accuracy a is % of reading and b is LSD (Least Significant Digit).
3. AC crest factor 2.5 at 4000 counts, And non-sinusoidal wave according to
wave factor calculated as follows:
a) Add 3% when crest factor is 1~2
b) Add 5% when crest factor is 2~2.5
Function
Range
Resolution
Accuracy
61-737
±(a%+b)
Continuity
0.1
Ω
≤
10
Ω
: Buzzer
beeps and red
indicator LED
illuminates
continuously
≥
70
Ω
: No buzzer
beep
Diode
0.001V
Silicon PN joint
with forward voltage
about 0.5V to 0.8V
Capacitance (F)
400.0nF
0.1nF
±(4.0%+8)
4.000uF
0.001uF
40.00uF
0.01uF
400.0uF
0.1uF
4.000mF
0.001mF
±(5.0%+9)
Frequency
(Hz/%)
10Hz to
9.999kHz
0.001Hz to
0.001kHz
±(0.1%+5)
Duty Cycle
(Hz/%)
0.1% to
99.9%
0.1%
±2.0%
Temp. °F
-40 to
1832°F
1°F
±(2.0%+6)
@ -40 to 32°F
±(2.0%+6) @33 to
1832°F
Temp. °C
-40 to
1000°C
1°C
±(2.0%+3)
@ -40 to 0°C
±(2.0%+3) @1 to
1000°C
NCV
40-600V
AC
≥
40V/(50 to 60Hz), with direct wire
contact, red indicator LED flashes at a
frequency of 3Hz, and the buzzer beeps at
a frequency of 3Hz