V DC Calibration 
61-700 (Refer to Figure 1)

 

1.   Set the function / range to 200mV DC. 
2.   Connect the calibrator to the 

V

 and

 COM

 inputs on the meter. 

3.   Output 190.0mV DC. 
      Adjust VR1 (VR 200

) until unit display reads 190.0 

      De-energize source and remove test leads 
 

61-701 (Refer to Figure 2A)

 

1.   Set the function / range to 20V DC. 
2.   Connect the calibrator to the 

V

 and

 COM

 inputs on the meter. 

3.   Output 19.00V DC. 
      Adjust VR1 (VR 200

) until unit display reads 19.00  

      De-energize source and remove test leads 
 

61-702 (Refer to Figure 3A) 
61-704 (Refer to Figure 4A)
 

1. 

Set the function / range to 2000mV DC. 

2. 

Connect the calibrator to the 

V

 and

 COM

 inputs on the meter. 

3. 

Output 1.900 VDC. 

Adjust VR1 (VR 200

) until unit display reads 1900. 

De-energize source and remove test leads 

 

V AC Calibration: 
61-704 only (Refer to Figure 4A) 

1.   Set the function/range to the 2000mV AC. 
2.   Connect the calibrator to the 

V

 and

 COM

 inputs on the meter. 

3.   Output 1.900 VAC/50Hz 

Adjust VR5 (VR 100

) until unit display reads 1900V± 1 digit. 

De-energize source and remove test leads

 

 

A AC Calibration: 
61-700 (Refer to Figure 1)
 
61-701 (Refer to Figure 2B; peel front label back to access calibration hole) 
61-702 (Refer to Figure 3B; peel front label back to access calibration hole) 
61-704 (Refer to Figure 4B; peel front label back to access calibration hole) 

1.      Set the function / range to the 200A AC 
2. 

Set output of the AC calibrator for 10.00A/60Hz +/- 0.01% and connect it to Coil = 10N = 100.0A 
AC. 

3. 

Clamp the jaws to the coil = 10N. 

4. 

Adjust VR2 (VR 2K

) for a display reading of 100.0 

±

1 digit 

 

Calibration of the 61-700 series is complete. 
Remove all leads from the calibrator and equipment.  
Return unit to proper operating condition. 

 
Form number TM61700-1-2-4 

Rev 4 September 2004 

Summary of Contents for 61-704

Page 1: ...mation provides the following information Precautions and safety information Specifications Performance test procedure Calibration and calibration adjustment procedure Basic maintenance replacing the...

Page 2: ...iance 2 Specifications 3 General Specification 3 Voltage Specifications 4 Current Specifications 4 Resistance Specifications 4 Frequency Specification 4 Capacitance Specifications 4 Diode Continuity S...

Page 3: ...al If you do not do so the protection provided by the meter may be impaired Read the Safety Information page before servicing this product In this manual a Warning identifies conditions and actions th...

Page 4: ...light will blink if there is Voltage Greater that 30V present at the input terminals o In addition to the LED and audible beep the 61 702 and 61 704 has a tactile vibration when voltage is present NCV...

Page 5: ...kaline 61 702 and 61 704 Low Battery Indicator symbol indicates low battery voltage Auto power off Approximately 60 minutes Dimensions 9 7 H X 3 7 W X 1 8 D 247mm H X 94mm W X 46mm D Maximum Cable Siz...

Page 6: ...ce 200 200K 1 0 3digit Diode Check 1mA 0 6mA Not specified Continuity 100 on Diode Continuity Not specified 61 702 61 704 Function Range Ranges Accuracy 2000mV 200V 50Hz 500Hz 1 2 3digits AC Voltage 2...

Page 7: ...74 1926 61 702 61 704 V AC 2000mV 1900mV AC 1kHz 1857 1943 61 702 61 704 V AC 200V 190V AC 50Hz 187 4 192 6 61 700 61 701 61 702 61 704 V AC 200V 190V AC 500Hz 187 4 192 6 61 700 61 701 61 702 61 704...

Page 8: ...alibrator to warm up Perform calibration at 23 2 C at relative humidity of 70 Temperature stabilization should be reached after 30 minutes 2 Disconnect the test leads and turn the range switch to OFF...

Page 9: ...V AC Calibration 61 704 only Refer to Figure 4A 1 Set the function range to the 2000mV AC 2 Connect the calibrator to the V and COM inputs on the meter 3 Output 1 900 VAC 50Hz Adjust VR5 VR 100 until...

Page 10: ...Page 8 Figure 1 Figure 2A Figure 2B Figure 3A Figure 3B Figure 4A Figure 4B Form number TM61700 1 2 4 Rev 4 September 2004...

Page 11: ...ps head screwdriver to remove the screws on battery cover 3 Remove battery from the battery compartment 4 Install new 9V battery NEDA 1604 An alkaline type is recommended 5 Install new battery into co...

Reviews: