Page 14
Testing the amp (A) Function
To verify the accuracy of AC current measurement functions, do the following:
1.Connect the Calibrator to the
A
and
COM
inputs on the Meter.
2. Turn the rotary switch to
A
.
3. Apply the inputs for steps 1-4 in Table 5.
4. For each input, compare the reading on the Meter display to the reading for your Meter in Table 5.
5. If the display reading falls outside of the range shown in Table 5, the meter does not meet specification.
Table 5 AC Current Test:
Step
Source
Frequency Reading
1
1.500A
50Hz
1.472A to 1.528A
2
1.500A
500Hz
1.472A to 1.528A
3
10.00A
50Hz
9.80A to 10.20A
4
10.00A
500Hz
9.80A to 10.20A
6. Turn the rotary switch to
A
.
7. Set the calibration for the current from step 1 - 2 in Table 6.
8. For each input, compare the reading on the Meter display to the reading in Table 6.
9. If the display reading falls outside of the range shown in Table 6, the meter does not meet specification.
Table 6 DC Current Test:
Step
Source
Reading
1
1.500A
1.483A to 1.517A
2
10.00A
9.88A to 10.12A
Testing the Frequency Function
To verify the accuracy of the Meter’s frequency function, do the following:
1. Connect the Calibrator to the
V
Ω
Hz
and
COM
inputs on the Meter.
Note:
The accuracy of the Calibrator’s frequency function must be appropriate for the specified accuracy of
the Meter.
2. Set the rotary switch to Hz.
3. Set the Calibrator for the sine wave voltage and frequency for steps 1-5 of Table 8.
4. Compare the reading on the Meter display with the display reading shown in Table 8.
5. If the display reading falls outside of the range shown in Table 8, the Meter does not meet specification.
Table 8 Frequency Test:
Step Soure
Level
Reading
1
1500Hz
100mV rms
1498Hz to 1502Hz
2
15.00KHz
100mV rms
14.98KHz to 15.02KHz
3
150.0KHz
100mV rms
149.8KHz to 150.2KHz
4
1.500MHz
250mV rms
1.498MHz to 1.502MHz
5
15.00MHz
1V rms
14.98MHz to 15.02MHz
*
Max. level: <5 Vac rms
Form number TM61354
Rev 2 July 2006