Page 5 

 
 
Placement of source for DC Balance Calibration 
 

 

 
 
Battery Replacement  (refer to Figure 2) 

1. Disconnect the test leads from any circuit under test and turn off meter. 
2. Use a Philips head screwdriver to remove the two screws on battery cover  
3. Remove battery from compartment and unsnap the battery connector. 
4. Install new 9V battery (NEDA #1604). An alkaline type is recommended.  
5. Install new battery into compartment using care not to pinch or bind battery leads 
6. Reinstall battery cover. 
 

 

Figure 2 

 
Form number TM61724 

Rev 2 Sept 2006 

Summary of Contents for 61-724

Page 1: ...des the following information Precautions and safety information Specifications Performance test procedure Calibration and calibration adjustment procedure Basic maintenance replacing the battery Form number TM61724 Revision 2 Date September 2006 Form number TM61724 Rev 2 Sept 2006 ...

Page 2: ...ons 2 General Specification 2 Voltage Specifications 2 Current Specifications 2 Resistance Specifications 2 Frequency Specification 2 Diode Specifications 2 Continuity Specifications 2 Performance Verification 3 Calibration 4 Clamp Jaw Alignment Figures 5 Replacing the Battery 5 Form number TM61724 Rev 2 Sept 2006 ...

Page 3: ...nly as specified CAUTION These statements identify conditions or practices that could result in damage to the equipment or other property WARNING These statements identify conditions or practices that could result in personal injury or loss of life Specific precautions Use proper Fuse To avoid fire hazard use only the fuse type and rating specified for this product Do not operate without covers To...

Page 4: ...plicable accuracy specification per C from 0 C to 18 C and 28 C to 50 C 32 F to 64 F and 82 F to 122 F Electrical Specification Accuracy are reading plus number of digits at 23 C 5 C 75 RH Function Setting Ranges Accuracy AC Voltage 200 0 600V at 50 60 Hz 1 2 4digits AC Voltage 200 0 600V at 40 500 Hz 2 0 4 digits DC Voltage 600V 0 5 1 digit AC Current 200 0 1000A at 50 60 Hz 1 75 4 digit AC Curre...

Page 5: ...he battery measures under 7 5V DC replace the battery before beginning the performance test 3 Input the values listed in Table 1 Table 1 Performance Verification Function Setting Range Input Low Limit High Limit ACV 200 ACV 200 190V AC 50Hz 190V AC 500Hz 187 3 185 8 192 7 194 2 ACV 600 ACV 600 500V AC 60Hz 500V AC 500Hz 490 486 510 514 DCV 600 500V DC 496 503 ACA 200A 100A 50Hz 97 9 102 1 ACA 200A...

Page 6: ...3 Output 100V DC Adjust R43 until unit display reads 100V De energize source and remove test leads DCA Zero Calibration 1 Remove the Jumper 2 Set the function range to the 200A DC 3 Without any signal applied keeping Jaw away from other devices adjust R54 for a display reading of 0 0 1 digit 4 Replace the Jumper then press the ZERO button adjust R506 for a display reading of 0 0 1 digit DCA Balanc...

Page 7: ...r 2 Use a Philips head screwdriver to remove the two screws on battery cover 3 Remove battery from compartment and unsnap the battery connector 4 Install new 9V battery NEDA 1604 An alkaline type is recommended 5 Install new battery into compartment using care not to pinch or bind battery leads 6 Reinstall battery cover Figure 2 Form number TM61724 Rev 2 Sept 2006 ...

Reviews: