IDEAL INDUSTRIES 61-362 Technical Manual Download Page 4

Page 2 

 
 

General specifications 

Model 61-362 
Characteristics  

Description 

Display  

3 ½ Digit LCD display 

Display Count 

3200 count, maximum reading 3199 
32 segment analog bar graph 

Over range Indication 

“OL” displayed 

Sampling Rate 

2 time/second 

Operating Environment: 
        Relative Humidity 

0

°

C to 50

°

C (32

°

F to 122

°

F) 70%RH 

Temperature Coefficient 

0.05X x (accuracy) per 

°

F (32

°

F to 65

°

F, 85

°

F to 122

°

Storage Environment: 

-20

°

C to 60

°

C (0

°

F to 140

°

F) at <70 relative humidity 

Power source: 

9V Battery (NEDA 1604) 

Battery Live: 

200 hours typical (alkaline) 

Low Battery Indicator: 

 symbol indicates low battery voltage 

Auto Power Off mode 

Approximately 25 minutes 
0.5A/250V fast acting fuse Type LA-3895 

A protection Fuse 

10A/600V fast acting fuse, Type LA-3897 

Dimensions 

5.75” H X 2.75” W X 1.5” D {without holster} 

Weight: 

Approximately 18.0 oz. including battery 

Safety 

UL1244, and Design to comply with IEC 1010-1 Cat III 

 
 

RANGES and ACCURACY SPECIFICATION 
61-362  
Function Setting 

Ranges 

Accuracy 

3.200V/32.00V/320.0V/ 600V 

AC Voltage 

   50Hz to 500Hz 

2.0% ± 4 digits 

320.0mV/32.00V/320.0V/600V 

1.2% ± 1digit 

DC Voltage 

3.200V 

0.8% ± 1digit 

320µA/3200 µA/32.00mA/320mA 

2.5% ± 4 digits 

 10A, 

AC Current 

   50Hz to 500Hz 

3.5% ± 4 digits 

320µA/3.20mA/32.00mA 2.0% 

 

± 1 digit

 

DC Current 

10A, 

3.0% ±  1 digit 

320.0

 

2.0% ± 3 digits 

3.200k/32.00k/320.0k 

1.5% ± 3 digits 

3.200M

 

2.5% ± 3 digits 

Resistance 

32.00M

 

5.0% ± 5 digits 

Continuity    

beep on < 100

 

Not Specified 

Diode Check  

 

2V DC max 

10% 

±

 2 digit 

 
AC Converter: 

  

Average responding, RMS Calibrated to Sine Wave

 

Overload Protection:

  

AC and DC Volts

: 600V DC or AC rms . 

 

 

 

Resistance, Diode, Continuity

: 500V DC or AC rms 

mA input:     

.5A/250V  DC/AC rms

 

Amps input:

 10A/600V DC/AC rms 

Form number 61362 

rev 1 May 2002 

Summary of Contents for 61-362

Page 1: ...ovides the following information Precautions and safety information Specifications Performance test procedure Calibration and calibration adjustment procedure Basic maintenance replacing the battery and fuses Form number 61362 Rev 1 date May 2001 Form number 61362 rev 1 May 2002 ...

Page 2: ... 1 Specifications 2 General Specification 2 Voltage Specifications 2 Current Specifications 2 Resistance Specifications 2 Diode Continuity Specifications 2 Performance Verification 3 Table 1 Verification 61 362 3 Calibration 4 Replacing the Battery 4 Replacing Fuses 5 Form number 61362 rev 1 May 2002 ...

Page 3: ... manual before starting Not all Caution Warning or Danger precautions are listed in this manual CAUTION These statements identify conditions or practices that could result in damage to the equipment or other property WARNING These statements identify conditions or practices that could result in personal injury or loss of life Specific precautions Use proper Fuse To avoid fire hazard use only the f...

Page 4: ...mensions 5 75 H X 2 75 W X 1 5 D without holster Weight Approximately 18 0 oz including battery Safety UL1244 and Design to comply with IEC 1010 1 Cat III RANGES and ACCURACY SPECIFICATION 61 362 Function Setting Ranges Accuracy 3 200V 32 00V 320 0V 600V AC Voltage 50Hz to 500Hz 2 0 4 digits 320 0mV 32 00V 320 0V 600V 1 2 1digit DC Voltage 3 200V 0 8 1digit 320µA 3200 µA 32 00mA 320mA 2 5 4 digits...

Page 5: ...9 36 30 64 ACV 32 30 00 500Hz 29 36 30 64 ACV 320 300V AC 50Hz 293 6 306 4 ACV 320 300V AC 500Hz 293 6 306 4 ACV 600 500V AC 50Hz 486 514 ACV 600 500V AC 60Hz 486 514 DCV 320mV 300mV DC 296 3 303 7 DCV 3 2V 3 000V DC 2 975 3 025 DCV 32V 30 00V DC 29 63 30 37 DCV 320 300 0V DC 296 3 303 7 DCV 600 500V DC 493 507 A DC 320µA 300µA 293 9 306 1 A DC 3200 µA 3000µA 2939 3061 A DC 32mA 30mA 29 39 30 61 A...

Page 6: ...lowing calibration procedure on an annual basis 61 362 Calibration Procedure Calibration 1 Set the Function Range Switch to the 3 200V DC position 2 Set the output of the DC calibrator for 3 000V and connect it to the V Ohm and COM input terminals 3 Adjust VR1 VR 200 ohm until the display reads 3 000V 1 digit Note This is the only adjustments required for the 61 362 Calibration is complete Battery...

Page 7: ...r in low resistance ohms mode to check the two fuses mA input 0 5A 250V fast acting fuse Amp input 10A 600V fast acting fuse 4 Remove the defective fuse with the recommended fuse mA with 0 5A 250V fast acting fuse Type LA 3895 is recommended Amp with 10A 600V fast acting fuse Type LA 3897 is recommended 5 Install bottom cover and secure with screws Form number 61362 rev 1 May 2002 ...